
 
We cannot reverse bias the IR photodiodes due 
to their high reverse saturation current. Increasing 
reverse bias increases the dark current and shot noise 
on the dark current would dominate other noise 
sources and reduce the SNR. Also, since the dark 
current increases with more reverse bias as the 
photocurrent remains constant, a smaller portion of 
the current flowing into the TIA is due to the 
photocurrent, reducing the SNR.  
2.3 Circuit Design 
Figure 5 shows the final circuit design used for each 
channel. We added the shot noise on the offset 
current, op amp input current noise, and op amp 
voltage noise for a number of op amps to select an 
op amp for the TIA. Here the offset current is the 
sum of the op amp input offset current and the 
current in R
T
 due to the op amp offset voltage. We 
included shot noise on the offset current since the 
source of the photodiode current is a diode junction. 
We chose the MAX4478 (Maxim, 2005) because it 
minimized the sum of these noise voltages. 
The chief factors affecting choice of the ADC 
are the resolution, number of delta-sigma (∆-Σ) 
blocks in the device, ability to daisy-chain serial data 
ports of devices, and conversion time. We chose a 
24-bit ∆-Σ ADC with 8 channels, 8 ∆-Σ  blocks, and 
maximum sampling rate of 52.7 kHz when using 
high resolution (Texas Instruments, 2008). Four 
ADCs are necessary to sample all 32 channels 
simultaneously. We need a minimum of 20 bit ADC 
resolution to keep the quantization error well below 
the system noise. Since there is one ∆-Σ block for 
each photodiode channel, multiplexing ADCs is 
unnecessary, allowing time efficiency and 
simplifying the software required to archive data.  
The serial ports of the ADC we chose can be 
daisy-chained which also simplifies the system 
because only one serial port is required. We need to 
record four types of samples to calculate the glucose 
concentration: with LED 1 on/LED2 off, both LEDs 
off, LED1 off/LED2 on, and both LEDs off. Since 
we sample all channels simultaneously and typically 
record N = 250 samples per second, recording four 
types of data samples requires a sampling rate of 1 
 
 
 
Figure 5: Schematic of one DAU channel. 
kHz, well below the ADS1278 maximum sample 
rate. We selected a low noise voltage reference with 
noise voltage of 1.5 μV and a low temperature 
coefficient of 0.6 PPM/deg. C. (Cirrus Logic, 2009) 
3 EXPERIMENT AND RESULTS 
Figure 6 shows the method we use to record 
experimental data. The time per data point, T
Data
, is 1 
second and N = 250 so that T
Sample
 = 1/250 second. 
The DAU samples the voltage on all channels at 
points labeled S
n
. At points labeled D
m
 the average 
value of the samples for all channels are calculated 
and archived. When we use the glucose sensor the 
four types of data mentioned above must be 
recorded in each sample period to calculate the 
glucose concentration. While verifying the SNR of 
the DAU system in this paper we utilize low noise 
resistors and only one of the four types of data is 
required since there are no LEDs present without the 
glucose sensor. 
To test the SNR we record a large (1k data 
points) set of experimental data and calculate the 
SNR. Figure 7 shows a plot of channel voltage vs. 
time for a representative channel. Figure 7 shows 
2000 data points from channel 1 taken at a rate of 1 
Hz – the last 1000 data points were used to calculate 
the SNR. The plot shows approximately 20 µV of 
drift with +/- 10 µV of noise in the last 1000 data 
points. The standard deviation of this portion of data 
is 9.12 μV and the SNR for channel 1 is calculated 
to be 40.4 dB. Fig. 8 shows the SNR for all 32 
channels. The mean SNR is 41.7 dB and the 
standard deviation of the SNR is 2.0 dB. Channels 
15 and 31 do have 5 to 7 dB more noise than the 
other channels. This is likely due to slight 
differences in noise characteristics of the op amps. 
Fig. 4 also shows the thermal noise present in the 
BIODEVICES 2010 - International Conference on Biomedical Electronics and Devices
198