loading
Papers Papers/2022 Papers Papers/2022

Research.Publish.Connect.

Paper

Paper Unlock

Authors: Dmitry Rogatkin and Oxana Smirnova

Affiliation: Moscow Regional Research and Clinical Institute “MONIKI” named after M.F.Vladimirskiy, Russian Federation

Keyword(s): Optics, Laser, Diagnostics, Results, Spectroscopy, Noninvasive, Medicine, Fiber, Measurements, Error, Uncertainty, Dispersion, Metrology.

Related Ontology Subjects/Areas/Topics: Biomedical Engineering ; Biomedical Equipment ; Biomedical Instruments and Devices ; Biomedical Metrology ; Spectroscopic Applications

Abstract: Over the last 10-15 years a large amount of methods and devices of noninvasive medical spectrophotometry including such techniques as in vivo Laser Fluorescent Diagnostics, Tissues Reflectance Oximetry, Laser Doppler Flowmetry, etc. has been developed and involved in a real clinical practice. In that number several problems of accuracy and reproducibility of clinical diagnostic results have been under discussion as well. But systematic metrological research in this field is still unknown. What dispersions and errors in diagnostic data can be estimated if measurements will be executed on the same object several times, by several doctors with different qualifications or using several devices from both the same and a different manufacturer? In this paper some results of the complex study of errors and uncertainties in diagnostic data caused by using an optical multi-fibers probe are presented. Dispersion and errors up to a level of +/-36,3% for the average registered values were discove red. It is shown that the interactive component of errors caused by interaction of the probe and a surveyed object gives the main contribution to the total uncertainty in diagnostic data. (More)

CC BY-NC-ND 4.0

Sign In Guest: Register as new SciTePress user now for free.

Sign In SciTePress user: please login.

PDF ImageMy Papers

You are not signed in, therefore limits apply to your IP address 3.145.156.250

In the current month:
Recent papers: 100 available of 100 total
2+ years older papers: 200 available of 200 total

Paper citation in several formats:
Rogatkin, D. and Smirnova, O. (2013). Optical Fiber Probe as a Source of Errors and Uncertainty in Measurements for Optical Noninvasive Diagnostic Devices and Techniques. In Proceedings of the International Conference on Biomedical Electronics and Devices (BIOSTEC 2013) - BIODEVICES; ISBN 978-989-8565-34-1; ISSN 2184-4305, SciTePress, pages 233-238. DOI: 10.5220/0004200002330238

@conference{biodevices13,
author={Dmitry Rogatkin. and Oxana Smirnova.},
title={Optical Fiber Probe as a Source of Errors and Uncertainty in Measurements for Optical Noninvasive Diagnostic Devices and Techniques},
booktitle={Proceedings of the International Conference on Biomedical Electronics and Devices (BIOSTEC 2013) - BIODEVICES},
year={2013},
pages={233-238},
publisher={SciTePress},
organization={INSTICC},
doi={10.5220/0004200002330238},
isbn={978-989-8565-34-1},
issn={2184-4305},
}

TY - CONF

JO - Proceedings of the International Conference on Biomedical Electronics and Devices (BIOSTEC 2013) - BIODEVICES
TI - Optical Fiber Probe as a Source of Errors and Uncertainty in Measurements for Optical Noninvasive Diagnostic Devices and Techniques
SN - 978-989-8565-34-1
IS - 2184-4305
AU - Rogatkin, D.
AU - Smirnova, O.
PY - 2013
SP - 233
EP - 238
DO - 10.5220/0004200002330238
PB - SciTePress