Authors: José Leonardo Ferreira ; Nadja Carvalho Pereira ; Marconi de Oliveria Júnior ; João Lucas da Silva ; Flávio Henrique Vasconcelos ; Carlos Julio Tierra-Criollo ; Raquel Britto and Verônica Franco Parreira

Affiliation: UFMG, Brazil

Keyword(s): Calibration, measurement uncertainty, respiratory pressures measuring, weighted least squares.

Related Ontology Subjects/Areas/Topics: Biomedical Engineering ; Biomedical Instruments and Devices ; Biomedical Metrology

Abstract: This article presents the results associated to calibration and evaluation of the measurement uncertainty of a digital respiratory pressures measuring system developed at the Biomedical Engineering Research & Development Laboratory (NEPEB) of the Department of Electrical Engineering – UFMG. The proposed method uses the least squares weighted regression to establish the measurement model and to evaluate the uncertainty. The standard expanded uncertainty estimated by the model was 1.4 kPa.

CC BY-NC-ND 4.0

Sign In Guest: Register as new SciTePress user now for free.

Sign In SciTePress user: please login.

PDF ImageMy Papers

You are not signed in, therefore limits apply to your IP address 3.133.148.76

In the current month:
Recent papers: 100 available of 100 total
2+ years older papers: 200 available of 200 total

Paper citation in several formats:
Leonardo Ferreira, J.; Carvalho Pereira, N.; de Oliveria Júnior, M.; Lucas da Silva, J.; Henrique Vasconcelos, F.; Julio Tierra-Criollo, C.; Britto, R. and Franco Parreira, V. (2008). APPLICATION OF WEIGHTED LEAST SQUARES TO CALIBRATE A DIGITAL SYSTEM FOR MEASURING THE RESPIRATORY PRESSURES. In Proceedings of the First International Conference on Biomedical Electronics and Devices (BIOSTEC 2008) - Volume 1: BIODEVICES; ISBN 978-989-8111- 17-3; ISSN 2184-4305, SciTePress, pages 220-223. DOI: 10.5220/0001052602200223

@conference{biodevices08,
author={José {Leonardo Ferreira}. and Nadja {Carvalho Pereira}. and Marconi {de Oliveria Júnior}. and João {Lucas da Silva}. and Flávio {Henrique Vasconcelos}. and Carlos {Julio Tierra{-}Criollo}. and Raquel Britto. and Verônica {Franco Parreira}.},
title={APPLICATION OF WEIGHTED LEAST SQUARES TO CALIBRATE A DIGITAL SYSTEM FOR MEASURING THE RESPIRATORY PRESSURES},
booktitle={Proceedings of the First International Conference on Biomedical Electronics and Devices (BIOSTEC 2008) - Volume 1: BIODEVICES},
year={2008},
pages={220-223},
publisher={SciTePress},
organization={INSTICC},
doi={10.5220/0001052602200223},
isbn={978-989-8111- 17-3},
issn={2184-4305},
}

TY - CONF

JO - Proceedings of the First International Conference on Biomedical Electronics and Devices (BIOSTEC 2008) - Volume 1: BIODEVICES
TI - APPLICATION OF WEIGHTED LEAST SQUARES TO CALIBRATE A DIGITAL SYSTEM FOR MEASURING THE RESPIRATORY PRESSURES
SN - 978-989-8111- 17-3
IS - 2184-4305
AU - Leonardo Ferreira, J.
AU - Carvalho Pereira, N.
AU - de Oliveria Júnior, M.
AU - Lucas da Silva, J.
AU - Henrique Vasconcelos, F.
AU - Julio Tierra-Criollo, C.
AU - Britto, R.
AU - Franco Parreira, V.
PY - 2008
SP - 220
EP - 223
DO - 10.5220/0001052602200223
PB - SciTePress