loading
Papers Papers/2022 Papers Papers/2022

Research.Publish.Connect.

Paper

Paper Unlock

Authors: Dmitry Mikhaylov 1 ; Timur Khabibullin 1 ; Igor Zhukov 1 ; Andrey Starikovskiy 1 ; Landish Gubaydulina 2 ; Natalya Romanchuk 2 and Vladimir Konev 1

Affiliations: 1 Engineering Centre of the National Research Nuclear University “MEPhI”, Russian Federation ; 2 The Clinical Hospital № 85, Russian Federation

Keyword(s): Electroactive Polymer, Endoscopic Capsule Retention System, Expandable Capsule, Angular Orientation, Cancer, Fluorescent Beacon.

Related Ontology Subjects/Areas/Topics: Biomedical Engineering ; Biomedical Instruments and Devices ; Wireless Systems

Abstract: The aim of the research is to propose the retention system of the wireless endoscopic capsule allowing more detailed observation of stomach, small intestine as well as other parts of the digestive tract. The capsule has an outer layer or whiskers made of an electroactive polymer material that expands under influence of an applied voltage. An electroactive polymer expands up to 50% of its original size and creates a roughly ball-shaped object approximately 3 cm in diameter (in case of whiskers – they change their angular orientation). The performance (changing the size) of the mock-up of the capsule with eight whiskers made of electroactive polymer was tested using a silicone model with sensitive sensors simulating the digestive tract of a person. The probability of failure-free operation and the mean time to failure for capsule and recording device integrated circuits were calculated.

CC BY-NC-ND 4.0

Sign In Guest: Register as new SciTePress user now for free.

Sign In SciTePress user: please login.

PDF ImageMy Papers

You are not signed in, therefore limits apply to your IP address 3.145.34.237

In the current month:
Recent papers: 100 available of 100 total
2+ years older papers: 200 available of 200 total

Paper citation in several formats:
Mikhaylov, D.; Khabibullin, T.; Zhukov, I.; Starikovskiy, A.; Gubaydulina, L.; Romanchuk, N. and Konev, V. (2014). Development of Retention System of the Autonomous Endoscopic Capsule and Its Functionalities. In Proceedings of the International Conference on Biomedical Electronics and Devices (BIOSTEC 2014) - BIODEVICES; ISBN 978-989-758-013-0; ISSN 2184-4305, SciTePress, pages 77-84. DOI: 10.5220/0004719400770084

@conference{biodevices14,
author={Dmitry Mikhaylov. and Timur Khabibullin. and Igor Zhukov. and Andrey Starikovskiy. and Landish Gubaydulina. and Natalya Romanchuk. and Vladimir Konev.},
title={Development of Retention System of the Autonomous Endoscopic Capsule and Its Functionalities},
booktitle={Proceedings of the International Conference on Biomedical Electronics and Devices (BIOSTEC 2014) - BIODEVICES},
year={2014},
pages={77-84},
publisher={SciTePress},
organization={INSTICC},
doi={10.5220/0004719400770084},
isbn={978-989-758-013-0},
issn={2184-4305},
}

TY - CONF

JO - Proceedings of the International Conference on Biomedical Electronics and Devices (BIOSTEC 2014) - BIODEVICES
TI - Development of Retention System of the Autonomous Endoscopic Capsule and Its Functionalities
SN - 978-989-758-013-0
IS - 2184-4305
AU - Mikhaylov, D.
AU - Khabibullin, T.
AU - Zhukov, I.
AU - Starikovskiy, A.
AU - Gubaydulina, L.
AU - Romanchuk, N.
AU - Konev, V.
PY - 2014
SP - 77
EP - 84
DO - 10.5220/0004719400770084
PB - SciTePress