APPLICATION OF WEIGHTED LEAST SQUARES TO CALIBRATE A DIGITAL SYSTEM FOR MEASURING THE RESPIRATORY PRESSURES
José Leonardo Ferreira, Nadja Carvalho Pereira, Marconi de Oliveria Júnior, João Lucas da Silva, Flávio Henrique Vasconcelos, Carlos Julio Tierra-Criollo, Raquel Britto, Verônica Franco Parreira
2008
Abstract
This article presents the results associated to calibration and evaluation of the measurement uncertainty of a digital respiratory pressures measuring system developed at the Biomedical Engineering Research & Development Laboratory (NEPEB) of the Department of Electrical Engineering – UFMG. The proposed method uses the least squares weighted regression to establish the measurement model and to evaluate the uncertainty. The standard expanded uncertainty estimated by the model was 1.4 kPa.
DownloadPaper Citation
in Harvard Style
Leonardo Ferreira J., Carvalho Pereira N., de Oliveria Júnior M., Lucas da Silva J., Henrique Vasconcelos F., Julio Tierra-Criollo C., Britto R. and Franco Parreira V. (2008). APPLICATION OF WEIGHTED LEAST SQUARES TO CALIBRATE A DIGITAL SYSTEM FOR MEASURING THE RESPIRATORY PRESSURES . In Proceedings of the First International Conference on Biomedical Electronics and Devices - Volume 1: BIODEVICES, (BIOSTEC 2008) ISBN 978-989-8111- 17-3, pages 220-223. DOI: 10.5220/0001052602200223
in Bibtex Style
@conference{biodevices08,
author={José Leonardo Ferreira and Nadja Carvalho Pereira and Marconi de Oliveria Júnior and João Lucas da Silva and Flávio Henrique Vasconcelos and Carlos Julio Tierra-Criollo and Raquel Britto and Verônica Franco Parreira},
title={APPLICATION OF WEIGHTED LEAST SQUARES TO CALIBRATE A DIGITAL SYSTEM FOR MEASURING THE RESPIRATORY PRESSURES},
booktitle={Proceedings of the First International Conference on Biomedical Electronics and Devices - Volume 1: BIODEVICES, (BIOSTEC 2008)},
year={2008},
pages={220-223},
publisher={SciTePress},
organization={INSTICC},
doi={10.5220/0001052602200223},
isbn={978-989-8111- 17-3},
}
in EndNote Style
TY - CONF
JO - Proceedings of the First International Conference on Biomedical Electronics and Devices - Volume 1: BIODEVICES, (BIOSTEC 2008)
TI - APPLICATION OF WEIGHTED LEAST SQUARES TO CALIBRATE A DIGITAL SYSTEM FOR MEASURING THE RESPIRATORY PRESSURES
SN - 978-989-8111- 17-3
AU - Leonardo Ferreira J.
AU - Carvalho Pereira N.
AU - de Oliveria Júnior M.
AU - Lucas da Silva J.
AU - Henrique Vasconcelos F.
AU - Julio Tierra-Criollo C.
AU - Britto R.
AU - Franco Parreira V.
PY - 2008
SP - 220
EP - 223
DO - 10.5220/0001052602200223