A generic ultrasound positioning system has
been described. Afterwards, the system has to be
adapted to the operating theater characteristics.
The system’s architecture is as follows:
Figure 1: System’s architecture.
Emitting transducers (at least 3 to determine
mobile’s position) need to be placed on the surgical
instrument on known position (in order to compute
the instrument’s position).
Receiving transducers have to be placed on the
ceiling to avoid non-line of sight issues. To ensure
good performance of the system, the coordinates of
receiving transducer should be known as much
accurate as possible.
The FPGA assigns a temporal interval to each
emitter module. It commands the emitter to generate
ultrasound chirps; simultaneously the ultrasonic time
of flight (TOF) to the receiving transducers will be
computed and stored (as well as ambient
temperature and humidity). After the FPGA sends
all surveyed data, the computer will obtain the
coordinates of all emitting transducers and the
instrument’s position.
2.1 Ambient Influence in Operating
Rooms
In this sub chapter, several factors that may affect
system’s performance will be analysed. This
analysis will take into account operating room (OR)
ambient characteristics and several others
considerations that may bring about measurement
inaccuracies.
Update rate: The frequency of the
measurements reported by the tracking system. The
higher this parameter the better tracking results are
obtained. The TOF system has a low update rate
caused by the low speed of sound and the sequential
triple emission of pulses.
Misalignment between transducer pairs: Due
to the quantity of receiving transducers and the wide
range of mobile’s movements, we cannot expect to
have perfect alignment between transducers pair
(emitter-receiver). It has been stated (Lamancusa,
1990) that misalignment between transducer pairs
will bring about a distance measurement error.
Ultrasound wave reflection on surfaces:
Reflections on surfaces has been used in ultrasound
distance measurement systems, using echoes to
determine distances to surfaces. Nevertheless,
considering the number of medical instruments in
OR (lamps, monitors, tables), echoes might bring
about measurement errors to our system and affect
system’s robustness. Therefore, reflection on
surfaces should be analyzed and taken into account
in system’s design.
Airflows: Most of these systems are based on
calculating the ultrasound TOF and multiplying it by
the sonic air speed to obtain the distance. Therefore,
airflow will modify the speed which the ultrasound
moves in the air.
Temperature: The temperature is an
environmental factor to consider when designing an
accurate ultrasound positioning system, as it affects
the speed of sound (Liao et al, 2004).
Relative Humidity: The relative humidity
modifies the speed of sound, affecting distance
measurement (Liao et al, 2004).
Table 1: Inaccuracies due to ambient factors.
Parameter Accuracy Max. Error
Temperature 0.3 º C 1.77 mm
Humidity 1.8 % RH 0.4 mm
Misalignment --- 1 mm
Airflows --- 1.3 mm
Total --- 4.4 mm
2.2 Distance Measurement
Once analysed the environmental factors (table 1)
that influence distance measurements, we will
analyse the measurement process. This part includes:
generation of ultrasonic chirps, reception and
treatment of the signals; and determination of TOFs.
TOF is the time elapsed between the
transmission of a pulse and its reception, from which
the target distance can be calculated multiplying
speed of sound in air by TOF. Using TOF to
measure the distance, the system errors are primarily
due to amplitude degradation of the received signal,
and uncertainty in the speed of sound.
There is a close relation between generation of
ultrasonic chirps and the method to determine TOFs
and distances (Tatar, 2006; Huang et al, 2002).
Moreover, the chirp generation method can affect
system’s performance. For example, some methods
require longer chirps, lowering system’s update rate
ULTRASONIC MOTION TRACKING OF INSTRUMENTS IN OPERATING THEATRE
187