LOW-COST ADAPTIVE METHOD FOR REAL-TIME ECG BASELINE WANDER REMOVAL WITH REDUCED P AND T WAVE DISTORTION
Rubén Fraile, Juan Ignacio Godino-Llorente
2009
Abstract
An adaptive algorithm for the removal of baseline wander in ECG is presented. The scheme is based on a single-tap LMS filter that estimates the baseline signal. The baseline is further processed by a moving average filter. This way, reduced distortion of P-Q and S-T segments is achieved with a low computational cost. Moreover, the proposed system has a short impulse response that makes it appropriate for real-time applications.
References
- Barati, Z. and Ayatollahi, A. (2006). Baseline wandering removal by using independent component analysis to single-channel ECG data. In Proc. of the International Conference on Biomedical and Pharmaceutical Engineering ICBPE 2006, pages 152-156. ID: 1.
- Laguna, P., Jane, R., and Caminal, P. (1992). Adaptive filtering of ECG baseline wander. In Proc. of the 14th Annual International Conference of the IEEE Engineering in Medicine and Biology Society, volume 14, pages 508-509.
- Physionet (2007). The MIT-BIH arrythmia database. URL: http://www.physionet.org/physiobank/database/mitdb/.
- S örnmo, L. (1991). Time-varying filtering for removal of baseline wander in exercise ECGs. In Proc. of Computers in Cardiology, pages 145-148.
- S örnmo, L. and Laguna, P. (2005). Bioelectrical Signal Processing in Cardiac and Neurological Applications. Elsevier Academic Press.
Paper Citation
in Harvard Style
Fraile R. and Godino-Llorente J. (2009). LOW-COST ADAPTIVE METHOD FOR REAL-TIME ECG BASELINE WANDER REMOVAL WITH REDUCED P AND T WAVE DISTORTION . In Proceedings of the International Conference on Bio-inspired Systems and Signal Processing - Volume 1: BIOSIGNALS, (BIOSTEC 2009) ISBN 978-989-8111-65-4, pages 476-479. DOI: 10.5220/0001534304760479
in Bibtex Style
@conference{biosignals09,
author={Rubén Fraile and Juan Ignacio Godino-Llorente},
title={LOW-COST ADAPTIVE METHOD FOR REAL-TIME ECG BASELINE WANDER REMOVAL WITH REDUCED P AND T WAVE DISTORTION},
booktitle={Proceedings of the International Conference on Bio-inspired Systems and Signal Processing - Volume 1: BIOSIGNALS, (BIOSTEC 2009)},
year={2009},
pages={476-479},
publisher={SciTePress},
organization={INSTICC},
doi={10.5220/0001534304760479},
isbn={978-989-8111-65-4},
}
in EndNote Style
TY - CONF
JO - Proceedings of the International Conference on Bio-inspired Systems and Signal Processing - Volume 1: BIOSIGNALS, (BIOSTEC 2009)
TI - LOW-COST ADAPTIVE METHOD FOR REAL-TIME ECG BASELINE WANDER REMOVAL WITH REDUCED P AND T WAVE DISTORTION
SN - 978-989-8111-65-4
AU - Fraile R.
AU - Godino-Llorente J.
PY - 2009
SP - 476
EP - 479
DO - 10.5220/0001534304760479