SYNCHRONIZING AN X-RAY AND ANESTHESIA MACHINE VENTILATOR - A Medical Device Interoperability Case Study
David Arney, Julian M. Goldman, Susan F. Whitehead, Insup Lee
2009
Abstract
When a x-ray image is needed during surgery, clinicians may stop the anesthesia machine ventilator while the exposure is made. If the ventilator is not restarted promptly, the patient may experience severe complications. This paper explores the interconnection of a ventilator and simulated x-ray into a prototype plug-and-play medical device system. This work assists ongoing interoperability framework development standards efforts to develop functional and non-functional requirements and illustrates the potential patient safety benefits of interoperable medical device systems by implementing a solution to a clinical use case requiring interoperability.
References
- Arney, D., Goldman, J., Lee, I., Llukacej, E., and Whitehead, S. (2007). Use Case Demonstration: X-Ray / Ventilator. In High Confidence Medical Devices, Software, and Systems and Medical Device Plug-and-Play Interoperability, 2007, page 160.
- ASTM F29 WK19878 (2008). New Specification for Equipment in the Integrated Clinical Environment - Part I: General Requirements for Integration.
- Langevin, P. B., Hellein, V., Harms, S. M., Tharp, W. K., Cheung-Seekit, C., and Lampotang, S. (1999). Synchronization of Radiograph Film Exposure with the Inspiratory Pause. Am. J. Respir. Crit. Care Med., 160(6):2067-2071.
- Lofsky, A. S. (2004). Turn Your Alarms On! Newsletter, 19(4):41-60.
- Nagle, J. (1984). Request for Comments: 896, Congestion Control in IP/TCP Internetworks. Technical report, Ford Aerospace and Communications Corporation.
- U.S. Department of Health and Human Services, Food and Drug Administration, Center for Drug Evaluation and Research (CDER), Center for Biologics Evaluation and Research (CBER) (2005). Guidance for Industry Development and Use of Risk Minimization Action Plans. Technical report, Office of Training and Communication, Division of Drug Information.
Paper Citation
in Harvard Style
Arney D., M. Goldman J., F. Whitehead S. and Lee I. (2009). SYNCHRONIZING AN X-RAY AND ANESTHESIA MACHINE VENTILATOR - A Medical Device Interoperability Case Study . In Proceedings of the International Conference on Biomedical Electronics and Devices - Volume 1: BIODEVICES, (BIOSTEC 2009) ISBN 978-989-8111- 64-7, pages 52-60. DOI: 10.5220/0001537100520060
in Bibtex Style
@conference{biodevices09,
author={David Arney and Julian M. Goldman and Susan F. Whitehead and Insup Lee},
title={SYNCHRONIZING AN X-RAY AND ANESTHESIA MACHINE VENTILATOR - A Medical Device Interoperability Case Study},
booktitle={Proceedings of the International Conference on Biomedical Electronics and Devices - Volume 1: BIODEVICES, (BIOSTEC 2009)},
year={2009},
pages={52-60},
publisher={SciTePress},
organization={INSTICC},
doi={10.5220/0001537100520060},
isbn={978-989-8111- 64-7},
}
in EndNote Style
TY - CONF
JO - Proceedings of the International Conference on Biomedical Electronics and Devices - Volume 1: BIODEVICES, (BIOSTEC 2009)
TI - SYNCHRONIZING AN X-RAY AND ANESTHESIA MACHINE VENTILATOR - A Medical Device Interoperability Case Study
SN - 978-989-8111- 64-7
AU - Arney D.
AU - M. Goldman J.
AU - F. Whitehead S.
AU - Lee I.
PY - 2009
SP - 52
EP - 60
DO - 10.5220/0001537100520060