ELECTRICAL CHARACTERIZATION OF SEQUENCE-SPECIFIC LABEL-FREE DNA BY USING POLYSILICON WIRE
You-Lin Wu, Po-Yen Hsu, Chung-Ping Hsu, Wen- Cheng Liu
2010
Abstract
In this paper we report the electrical measurement of sequence-specific and label-free deoxyribonucleic acid (DNA) by using poly-silicon (poly-Si) wire. Four single-strained (ss) DNA bases, adenine (A), thymine (T), cytosine (C) and guanines (G), as well as double-stranded (ds) DNA sequences, ds(A-T) and ds(C-G), with different lengths and concentrations were dropped onto the poly-Si wire surface, and currents flowing through the poly-Si wire channel were determined. It is found that the amount change of the channel current I for ssDNAs with fixed length and fixed concentration is that T > C > G > A. For dsDNA sequence, we observe that ds(A-T) has higher I than does ds(C-G). We also prove that single base change in ssDNA is feasible by using the poly-Si wire sensor.
References
- de-los-Snatos-Alvarez, P., Lobo-Castanon M. Juses, Miranda-Ordieres, A. J., Tunon-Blanco, (2004). Current strategies for electrochemical detection of DNA with solid electrodes. Anal. Bioanal.Chem., 378, 104-118.
- Fritz, J., Cooper, E. B., Gaudet, S., Sorger, P. K. Manalis, S. R. (2002). Electronic detection of DNA by its intrinsic molecular charge. Proc. Natl. Acad. Sci. U.S.A., 99, 14142-14146.
- Hahm, J. I., & Lieber, C. M. (2001). Direct ultrasensitive electrical detection of DNA and DNA sequence variations using nanowire nanosensors. Nano lett., 4, 51-54.
- Kim, K. S., Ahn, S. K., Lee, Y., Lee, J. M., Roh, Y. (2006). Effects of gate voltage on the characteristics of sourcedrain current formed through DNA molecules. Thin Solid Films, 515, 822-826.
- Lander, E. S. (1999). Array of hope. Nat Genet. 21, 3-4.
- Mastrangelo, C. H., (1999). DNA analysis system on a chip. Solid-State Chemical and Biochemical Sensors, Techna Faenza., 495-476.
- Storm, A. J., van Noort, J., de Vries, S. Dekker, C. (2001). Insulating behaviour for DNA molecules between nanoelectrodes at the 100nm length scale. Appl. Phys. Lett., 79, 3881.
- Wang, J. (2000). Towards Genoelectronics: Electrochemical Biosensing of DNA Hybridization Chem. Eur. J., 5, 1681-1685.
- Wu, Y. L., Hsu, P. Y., Hu, S. H., Hung, W. C., Lin, S. T., Hsu, C. P., Cullis, A. G. (2008). Polysilicon Wire Sensor for Biochemical Detection with The Help of Capillary Atomic-Force-Microscopy Tip for Solution Transfer. 9th International conference on Solid State and Integrated Circuit technology. IEEE press.
- Xu, M., Endres, R. Arakawa, G., Y. (2007). The Electronic Properties of DNA bases, Small, 3, 1539- 1543.
- Zwolak, M., Di Ventra, M., (2008). Colloquium: Physical approaches to DNA sequencing and detection. Rev. Mod. Phys., 80, 141-165.
Paper Citation
in Harvard Style
Wu Y., Hsu P., Hsu C. and Liu W. (2010). ELECTRICAL CHARACTERIZATION OF SEQUENCE-SPECIFIC LABEL-FREE DNA BY USING POLYSILICON WIRE . In Proceedings of the Third International Conference on Biomedical Electronics and Devices - Volume 1: BIODEVICES, (BIOSTEC 2010) ISBN 978-989-674-017-7, pages 170-173. DOI: 10.5220/0002712601700173
in Bibtex Style
@conference{biodevices10,
author={You-Lin Wu and Po-Yen Hsu and Chung-Ping Hsu and Wen- Cheng Liu},
title={ELECTRICAL CHARACTERIZATION OF SEQUENCE-SPECIFIC LABEL-FREE DNA BY USING POLYSILICON WIRE},
booktitle={Proceedings of the Third International Conference on Biomedical Electronics and Devices - Volume 1: BIODEVICES, (BIOSTEC 2010)},
year={2010},
pages={170-173},
publisher={SciTePress},
organization={INSTICC},
doi={10.5220/0002712601700173},
isbn={978-989-674-017-7},
}
in EndNote Style
TY - CONF
JO - Proceedings of the Third International Conference on Biomedical Electronics and Devices - Volume 1: BIODEVICES, (BIOSTEC 2010)
TI - ELECTRICAL CHARACTERIZATION OF SEQUENCE-SPECIFIC LABEL-FREE DNA BY USING POLYSILICON WIRE
SN - 978-989-674-017-7
AU - Wu Y.
AU - Hsu P.
AU - Hsu C.
AU - Liu W.
PY - 2010
SP - 170
EP - 173
DO - 10.5220/0002712601700173