2 STANDARD IMPEDANCE
MEASUREMENT
The impedance was measured using a modified FOT
setup, able to assess the respiratory mechanics from
4-50 Hz. The specifications of the device are: 11kg,
50x50x60 cm, 40 seconds measurement time,
European Directive 93/42 on Medical devices and
safety standards EN60601-1.
Typically for lung function testing purposes, the
subject is connected to the setup from figure 1 via a
mouthpiece, suitably designed to avoid flow leakage
at the mouth and dental resistance artefact. The
oscillatory pressure is generated by a loudspeaker
(LS), which is connected to a chamber. The LS is
driven by a power amplifier, which is fed with the
oscillatory signal generated by a computer (denoted
by U in figure 1-A and by U
g
in figure 1-B). The
movement of the LS cone generates a pressure
oscillation inside the chamber, which is applied to
the patient's respiratory system by means of a tube
connecting the LS chamber and the bacterial filter
(bf). A side opening of the main tubing (BT) allows
the patient to have fresh air circulation. Ideally, this
BT pipeline will have high impedance at the
excitation frequencies to avoid the loss of power
from the LS pressure chamber. It is advisable that
during the measurements, the patient wears a nose
clip and keeps the cheeks firmly supported. Before
starting the measurements, the frequency response of
the transducers (PT for pressure measurement) and
of the pneumotachograph (PN for flow
measurement) is calibrated.
Figure 1: A schematic overview (A) and an electrical
analogy of the FOT setup (B).
The measurements of air-pressure P and air-flow
Q during the FOT lung function test are done at the
mouth of the patient. The FOT excitation signal was
kept within a peak-to-peak range of 0.1-0.3 kPa, in
order to ensure optimality, patient comfort and
linearity (Oostveen et al., 2003). From these signals,
the non-parametric representation of the patient’s
lung impedance Z
r
is obtained assuming a linear
dependence between the breathing and
superimposed oscillations at the mouth of the patient
(Daroczy and Hantos, 1982; Ionescu and De Keyser,
2003).
Consider the equivalent circuit for the global
setup, denoted by figure 1-B, with the notations as:
U
g
= generator test signal – driving signal
(measured); U
r
= effect of spontaneous breathing
(respiratory system / unknown); Z
r
= impedance of
interest (to be estimated): the impedance of the total
respiratory system (including the airways, lung
tissues and chest wall); Z
1
= impedance (unknown)
describing the transformation of driving voltage (U
g
)
to chamber pressure; Z
2
= impedance (unknown) of
both bias tubes and loud-speaker chamber; Z
3
=
impedance (unknown) of tube segment between bias
tube and mouth piece (effect of pneumotachograph
essentially); P = (measured) pressure; Q =
(measured) flow. The corresponding equation is:
() () () ()
rr
sZsQsUs=+
(1)
where s denotes the Laplace operator. Since the
excitation signal is designed such that it is not
correlated with the breathing of the patient,
correlation analysis can be applied to the measured
signals. Therefore, one can estimate the respiratory
impedance as the ratio:
()
()
()
g
g
PU
r
QU
Sj
Zj
Sj
ω
ω
ω
=
(2)
where the P corresponds to pressure (its electrical
equivalent is voltage) and Q corresponds to air-flow
(its electrical equivalent is current), U
g
the excitation
signal,
()
ij
Sj
ω
the cross-correlation spectra
between the various input-output signals, ω is the
angular frequency and
1/2
(1)j =− , resulting in the
complex variable Z
r
. From the point of view of the
forced oscillatory experiment, the signal components
of respiratory origin (U
r
) have to be regarded as pure
noise for the identification task (Ljung, 1999).
3 THEORETICAL BASIS
The following input-output relationship can be
written based on figure 1-B (Ionescu and De Keyser,
2003):
LS
PT
bf
subject
P(t)Q(t)
LS
BT
PT
PN
bf
subject
DAQ board
Laptop / GUI
U(t)
P(t) Q(t)
LS
PT
bf
subject
P(t)Q(t)
LS
BT
PT
PN
bf
subject
DAQ board
Laptop / GUI
U(t)
P(t) Q(t)
A
B
Z
2
Z
1
Z
3
Z
r
P
Q
U
r
U
g
Z
2
Z
1
Z
3
Z
r
P
Q
U
r
U
g
Z
FOT
LS
PT
bf
subject
P(t)Q(t)
LS
BT
PT
PN
bf
subject
DAQ board
Laptop / GUI
U(t)
P(t) Q(t)
LS
PT
bf
subject
P(t)Q(t)
LS
BT
PT
PN
bf
subject
DAQ board
Laptop / GUI
U(t)
P(t) Q(t)
A
B
Z
2
Z
1
Z
3
Z
r
P
Q
U
r
U
g
Z
2
Z
1
Z
3
Z
r
P
Q
U
r
U
g
Z
FOT
Z
2
Z
1
Z
3
Z
r
P
Q
U
r
U
g
Z
2
Z
1
Z
3
Z
r
P
Q
U
r
U
g
Z
FOT
BIODEVICES 2012 - International Conference on Biomedical Electronics and Devices
120