NON-LINEAR ANALYSIS OF FETAL HEART RATE IN CARDIOTOCOGRAPHY USING SAMPLE ENTROPY
J. A. L. Marques, P. C. Cortez, J. P. V. Madeiro, F. S. Schlindwein
2012
Abstract
The complex system of mother and foetus interacting during pregnancy contains both dependent and independent subsystems and it is unlikely that it can be studied using only linear techniques. Considering this, the conventional medical analysis of Fetal Heart Rate (FHR) based on Cardiotocography (CTG) traces can be expanded considering nonlinear approaches. This work presents the use of Sample Entropy (SampEn) as a measure of system complexity, using a 5 minutes window of FHR signal (1200 samples), using values for parameters m and r based on literature to analyse the signal complexity behaviour in time. The database is comprised of 22 pre-classified intrapartum exams, expected to have a high degree of time domain dynamics. The analysis shows that severe FHR decelerations result in small values of SampEn, reflecting a low level of complexity. On the other hand, a set of high level transient FHR accelerations also causes the same effect. The occurrences of repetitive patterns (similar to sinusoidal waves, which are pathological) cause a drop of SampEn values. The results encourage us to consider SampEn as one viable parameter for nonlinear FHR signal analysis.
References
- Doganaksoy, A., & Göloglu, F. (2009). On Lempel-Ziv Complexity of Sequences. Athenas: Department of Mathematics.
- Ingemarsson, I., Ingemarsson, E., & Spencer, J. A. (1993). Fetal Heart Rate Monitoring - A Practical Guide. New York: Oxford Medical Publications - Oxford University Press.
- Kaplan, D., & Staffin, P. (2008). Retrieved November 30, 2009, from Software for Heart Rate Variability: http://www.macalester.edu/kaplan/hrv/doc/
- Marques, J. A., Cortez, P. C., & Madeiro, J. P. (2010). Fetal Heart Rate Variability Analysis in Cardiotocographic Exams. Brazilian Conference on Health Informatics. Porto de Galinhas: SBIS.
- Mathworks. (2009). Matlab. Retrieved Julho 10, 2011, from Mathworks: http://www.mathworks.com
- Pincus, S. M. (1991). Approximate Entropy as a measure of System Complexity. Proceedings of the National Academy of Science , pp. 2083-2088.
- Richman, J. S., & Moorman, J. R. (2000). Physiological time-series analysis using approximate entropy and sample entropy. American Jounal of Physiology - Heart and Circulatory Physiology , pp. H2039-H2049.
- Kantz, H., & Schreiber, T. (1997). Nonlinear Time Series Analysis. Cambridge. Cambridge University Press.
Paper Citation
in Harvard Style
A. L. Marques J., Cortez P., Madeiro J. and Schlindwein F. (2012). NON-LINEAR ANALYSIS OF FETAL HEART RATE IN CARDIOTOCOGRAPHY USING SAMPLE ENTROPY . In Proceedings of the International Conference on Bio-inspired Systems and Signal Processing - Volume 1: BIOSIGNALS, (BIOSTEC 2012) ISBN 978-989-8425-89-8, pages 244-248. DOI: 10.5220/0003776502440248
in Bibtex Style
@conference{biosignals12,
author={J. A. L. Marques and P. C. Cortez and J. P. V. Madeiro and F. S. Schlindwein},
title={NON-LINEAR ANALYSIS OF FETAL HEART RATE IN CARDIOTOCOGRAPHY USING SAMPLE ENTROPY},
booktitle={Proceedings of the International Conference on Bio-inspired Systems and Signal Processing - Volume 1: BIOSIGNALS, (BIOSTEC 2012)},
year={2012},
pages={244-248},
publisher={SciTePress},
organization={INSTICC},
doi={10.5220/0003776502440248},
isbn={978-989-8425-89-8},
}
in EndNote Style
TY - CONF
JO - Proceedings of the International Conference on Bio-inspired Systems and Signal Processing - Volume 1: BIOSIGNALS, (BIOSTEC 2012)
TI - NON-LINEAR ANALYSIS OF FETAL HEART RATE IN CARDIOTOCOGRAPHY USING SAMPLE ENTROPY
SN - 978-989-8425-89-8
AU - A. L. Marques J.
AU - Cortez P.
AU - Madeiro J.
AU - Schlindwein F.
PY - 2012
SP - 244
EP - 248
DO - 10.5220/0003776502440248