CT scanner X-ray beam leads to substantial
distortions in dose estimates despite the fact that the
last method takes into account beam rotation
.TLDs
(LiF:Mg, Ti; 4.5±0.12 mm in diameter, 0.9±0.1 mm
thickness) from chosen group were placed inside a
container in which 22 – 24 sensors are housed
simultaneously. As a reference dosimeter Unfors Xi
with pencil type ionization chamber has been used.
Container with TLDs as well as Unfors Xi detector
were installed at the central and periphery holes of
CTDI phantom located in the centre of scanner
gantry and after this several exposures for one tube
rotation and with fixed table position have been
made. Multiple exposures are needed in order to
reliably exceed the sensitive threshold of
thermoluminescent dosimeters. Considering
reference device (Unfors Xi) readings and TLDs
measurements (estimates of an integral under
absorbed dose distribution curve on ±50 mm
interval) two values of the weighted CTDI have
been calculated and their ratio gives the calibration
factor for testing TLDs group.
Effective doses on DLP basis have been
calculated in respect that conversion factor for adult
chest studies in accordance with appropriate
document is equal to 0.017 mSv·mGy
-1
·cm
-1
(European Commission’s Study Group, 1999).
An adult (hermaphrodite) anthropomorphic body
phantom consists of 25 mm elements simulating
human body organ sections with different densities.
This phantom with TLDs (about 100 sensors) placed
inside was chosen for reference measurements. One
phantom section with thermoluminescent sensors
housed in is shown on Fig.1, as an example. TLDs
distribution in phantom was the following: red bone
marrow – 7; colon – 5; lungs – 24; stomach – 4;
breast – 2; gonads – 2; bladder – 2; liver – 5;
esophagus – 8; thyroid – 6; skin – 11; bone surface –
10; remind organs – 13. Using an anthropomorphic
phantom+TLDs, effective dose evaluation has been
made on the basis of weighting factors for different
organs and tissue presented in International
Commission on Radiological Protection (ICRP)
documents: ICRP Publication 60 and ICRP
Publication 103 recommendations (ICRP, 1991);
(ICRP, 2007).
3 RESULTS AND DISCUSSION
The right parts of absorbed dose distribution curves
received at the central and periphery holes of
doubled CTDI “body” phantom are presented on
Fig.2 and Fig.3 respectively. Fig.4 shows the right
parts of weighted distribution curves. Full length
weighted curves have been used for weighted CTDI
calculation on the basis of integrals under the curves
on the interval from –200 to +200 mm). Weighting
has been made using weighted factors 1/3 and 2/3
for centre and periphery zones respectively. It
follows from the received data that the integrals
under the absorbed dose distribution curves at the
central hole of doubled CTDI phantom (in the range
of ±200 mm) differ from the integrals on the interval
from –50 to +50 mm (traditional CTDI method)
approximately by (53 – 65)% for both scanners. For
periphery holes this difference is not so significant
and is about (11 – 13)%. Mentioned above means
that ignoring of scattered radiation outside ±50 mm
zone along z-axis leads to underestimation of the
effective dose up to approximately 20% in average
when using the traditional method on the basis of
single CTDI "body" phantom and 100 mm
ionization chamber (weighted CTDI) in case of
collimation 32 – 40 mm.
Figure 1: Anthropomorphic phantom section with TLDs
housed in.
Figure 2: The right part of absorbed dose distribution at
the centre hole of doubled CTDI phantom.
BIODEVICES2013-InternationalConferenceonBiomedicalElectronicsandDevices
92