2.2 Other Analysis Techniques
There are other techniques that are also suitable for
the volatile organic compounds analysis, like GC-
MS, SIFT-MS and PTR-MS.
GC-MS (mass spectrometry with gas
chromatography) is a technique that enables the
separation and identification of volatile organic
compounds and some volatile inorganic compounds
of a gas mixture (Dolan, Newman and Stauffer,
2007). This technique needs a long time to perform
the analysis and pre-concentration (Blake, Monks
and Ellis, 2009).
SIFT-MS (mass spectrometry associated with a
selected ion flow tube) is a technique that uses
precursor ions to ionize gases in a gas sample. It is a
technique that cannot be miniaturized and it is less
sensitive than the MCC-IMS (Spanel and Smith,
1996).
PTR-MS (mass spectrometry by proton transfer
reaction) is a technique that enables the
identification of volatile organic compounds mostly
from natural sources. It uses a vacuum system and
for that reason it cannot be miniaturized (Blake et.
al., 2009).
Compared with other methods of VOC analysis,
ion mobility spectrometry (IMS) stands out due to
its high sensitivity, low cost, portability and
simplicity. The fact that is does not require vacuum
or further sample preparation and the analysis is
performed in a few minutes makes this technique
suitable to be used in hospitals and healthcare
centers.
2.3 Materials and Methods
The samples were collected from 9 diabetic patients
and 9 health individuals according to the standard
protocol procedures with sterile tear flow test strips
(by Sno*Strips) based on the technique of the
Schirmer’s Test (Zhou and Beuerman, 2012). In this
test, the sterile strip is placed in the outer lower
eyelid of each eye during 3 minutes. After this time,
the strip is collected and stored into vials which are
closed with screw caps with a silicone septum.
The reason why the strip is placed in a specific
part of the eyelid is related to some facts:
The lacrimal gland which is responsible for the
production of the aqueous layer of the lacrimal fluid
is located at the superior temporal portion of each
eye,
The lacrimal channels are located at the nasal
portion of each eye and communicate with the
lacrimal sac, which could cause a contamination of
the lacrimal sample with nasal compounds.
The cornea is an extremely sensitive eye region
that could be negatively affected by the strip contact.
A strip with collected lacrimal fluid was placed
in a 20 mL vial, sealed and heated in an Analogic
Heating Plate from VWR®, at 60ºC during 10
minutes. The vial was connected to the MCC-IMS
through a needle that was inserted across the silicone
septum. Figure 3 shows the experimental scheme
considered in this study
After the 10 min of equilibrium headspace the
carrier gas (N
2
, 25 mL min−1) transferred the
injected sample to the MCC for separation. Then,
the separated analyte, was driven into the ionization
chamber of the ion mobility spectrometer.
Figure 3: Experimental scheme considered in this study.
Analyses were performed on a MCC-IMS
apparatus fabricated by Gesellschaft für Analytische
Sensorsysteme (G.A.S. mbH, Dortmund, Germany).
The multicapillary column (Multichrom, Ltd.,
Novosibirsk, Russia) has a length of 20 cm, a
volume of 0.45 mL, providing a high sample
capacity for preseparation.. The detector was
equipped with a Tritium ioniztion asource (St.
Petersburg, Russia) with an activity of 300 MBq. A
sample inlet lets a continuous stream of nitrogen 5.0
(Air Liquide, Portugal) at 25 mL min−1 to pass
through the ionization chamber where ions are
formed and focused to a shutter grid. A drift gas
flow rate of 500 mL min−1 was used to provide a
good separation of ions and to reduce the flush time
between consecutive measurements.
All experimental parameters of MCC-IMS , such
as drift gas and carrier gas flow rates, injection
volume, grid pulse and system temperature, were
optimized in order to obtain the better spectra.
BIODEVICES2013-InternationalConferenceonBiomedicalElectronicsandDevices
222