Abstract

In er we present the desigthis papn of a wearable temperature sensing device for remote neonatal monitoring. It is designed for continuous and real-time monitoring of the infants in remote rural areas, for the first few weeks after their birth. It is capable of sensing the neonate’s skin temperature with 0.1◦ C accuracy to detect the early onset of hypothermia. The sensed data is transferred securely over bluetooth low energy radio to a nearby gateway, which then relays the information to a central database for real time monitoring. The device incorporates a medical grade thermistor which is directly interfaced to a microcontroller with an integrated bluetooth low energy radio. Low power optimizations at both the circuit and software levels ensure sleep currents of only 1uA, ensuring very long battery life. The device is packaged in a baby friendly, water proof housing and is easily sterilizable and reusable.

References

  1. Alf Helge, O. (2010). Bluetooth Low Energy: Wireless Connectivity for Medical Monitoring. Journal of Diabetes Science and Technology, 4(2).
  2. Artem, D., Steve, H., Stuart, T., and Joshua, S. (2013). Power consumption analysis of Bluetooth Low Energy, Zigbee, and ANT sensor nodes in cyclic sleep scenario. IEEE International Wireless Symposium, Beijing, China.
  3. Bluetooth (2012). Health tempertute profile. https:// developer.bluetooth.org/TechnologyOverview/Pages/ HTP.aspx.
  4. Boano, C., Lasagni, M., Romer, K., and Lange, T. (2011). Accurate temperature measurements for medical research using body sensor networks. In Object/Component/Service-Oriented Real-Time Distributed Computing Workshops (ISORCW), 2011 14th IEEE International Symposium on, pages 189-198.
  5. Cantherm (2009). High precision ntc thermistors. http:// www.cantherm.com/products/thermistors/mf51e.html.
  6. Chen, W., Dols, S., Oetomo, S. B., and Feijs, L. (2010). Monitoring body temperature of newborn infants at neonatal intensive care units using wearable sensors. In Proceedings of the Fifth International Conference on Body Area Networks, BodyNets 7810, pages 188- 194, New York, NY, USA. ACM.
  7. Embrace (2012). A low cost infant warmer. embraceglobal.org/.
  8. Global Health (2010). Causes of neonatal and child mortality in india: a nationally representative mortality survey. The Lancet, 376(9755):1853 - 1860.
  9. Gomez, C., Oller, J., and Paradells, J. (2012). Overview and evaluation of bluetooth low energy: An emerging lowpower wireless technology. Sensors, 12(9):11734- 11753.
  10. Hawley E, M.-D., Salvatore P, I., and Jonathan A, B. (2012). Movement toward a novel activity monitoring device. Sleep Breath, 16(3):913-917.
  11. Isetta, V., Agustina, L., Lopez Bernal, E., Amat, M., Vila, M., Valls, C., Navajas, D., and Farre, R. (2013). CostEffectiveness of a New Internet-Based Monitoring Tool for Neonatal Post-Discharge Home Care. Journel of Medical Internet Research, (15(2):e38).
  12. iThermometer (2012). Bluetooth digital thermometer for android devices. http://download.chinavasion.com/ download/CVXX-A140.pdf.
  13. Kumar, V., Mohanty, S., and Kumar, A. (2008). Effect of community-based behaviour change management on neonatal mortality in shivgarh, uttar pradesh, india: a cluster-randomised controlled trial. The Lancet, 372(9644):1151 - 1162.
  14. Kumar, V., Shearer, J. C., Kumar, A., and Darmstadt, G. L. (2009). Neonatal hypothermia in low resource settings: a review. Journal of Perinatology, 29:401-412.
  15. Macfarlane, F. (2006). Paediatric anatomy, physiology and the basics of paediatric anaesthesia.
  16. Neurosynaptics (2002). http://www.neurosynaptic.com/.
  17. Oestergaard, M. Z. and Inoue, M. e. a. (2011). Neonatal mortality levels for 193 countries in 2009 with trends since 1990: A systematic analysis of progress, projections, and priorities. PLoS Med, 8(8):e1001080.
  18. Rutter, N. (2000). Clinical consequences of an immature barrier. Seminars in Neonatology, 5(4):281 - 287.
  19. Scavacini, A. S. A.-l., Miyoshi, M. H., Kopelman, B. I., and Peres, C. A. d. A. A. (2007). Chest Expansion for Assessing Tidal Volume in Premature Newborn Infants on Ventilators. Jornal de Pediatria, 83:329 - 334.
  20. Stratasys (2013). High-quality, precise, multimaterial prototypes in a compact system. http://www.stratasys.com/3d-printers/design-series/ precision/objet260-connex.
  21. Susan, B., Debra, D., and Lori A, L. (2001). Neonatal Thermal Care, Part II: Microbial Growth Under Temperature Probe Covers. Journal Neonatal Network: The Journal of Neonatal Nursing, 20:19-23.
  22. TI (2008). 2.4ghz inverted f- antenna. http://www.ti.com/ lit/an/swru120b/swru120b.pdf.
  23. WHO (1993). Thermal Control of the Newborn: a Practical Guide. Maternal and Safe Motherhood Programme.
  24. WHO (2012). Newborn deaths decrease but account for higher share of global child deaths.
  25. WHO, UNICEF, and UN (2012). Levels and Trends in Child Mortality Report.
Download


Paper Citation


in Harvard Style

Rao H., Saxena D., Kumar S., G. V. S., Amrutur B., Mony P., Thankachan P., Shankar K., Rao S. and Bhat S. (2014). Low Power Remote Neonatal Temperature Monitoring Device . In Proceedings of the International Conference on Biomedical Electronics and Devices - Volume 1: BIODEVICES, (BIOSTEC 2014) ISBN 978-989-758-013-0, pages 28-38. DOI: 10.5220/0004798300280038


in Bibtex Style

@conference{biodevices14,
author={Hiteshwar Rao and Dhruv Saxena and Saurabh Kumar and Sagar G. V. and Bharadwaj Amrutur and Prem Mony and Prashanth Thankachan and Kiruba Shankar and Suman Rao and Swarnarekha Bhat},
title={Low Power Remote Neonatal Temperature Monitoring Device},
booktitle={Proceedings of the International Conference on Biomedical Electronics and Devices - Volume 1: BIODEVICES, (BIOSTEC 2014)},
year={2014},
pages={28-38},
publisher={SciTePress},
organization={INSTICC},
doi={10.5220/0004798300280038},
isbn={978-989-758-013-0},
}


in EndNote Style

TY - CONF
JO - Proceedings of the International Conference on Biomedical Electronics and Devices - Volume 1: BIODEVICES, (BIOSTEC 2014)
TI - Low Power Remote Neonatal Temperature Monitoring Device
SN - 978-989-758-013-0
AU - Rao H.
AU - Saxena D.
AU - Kumar S.
AU - G. V. S.
AU - Amrutur B.
AU - Mony P.
AU - Thankachan P.
AU - Shankar K.
AU - Rao S.
AU - Bhat S.
PY - 2014
SP - 28
EP - 38
DO - 10.5220/0004798300280038