Biodegradable Passive Resonance Sensor
Fabrication and Initial Testing
Timo Salpavaara
1
, Ville Ellä
2
, Minna Kellomäki
2
and Jukka Lekkala
1
1
Department of Automation Science and Engineering, Tampere University of Technology,
Korkeakoulunkatu 3, Tampere, Finland
2
Department of Electronics and Communications Engineering, Tampere University of Technology, Tampere, Finland
Keywords: Biodegradable, Passive Resonance Sensor.
Abstract: Biodegradable resonance circuits were studied. The circuits have a novel two-layer resonator structure
without galvanic through hole vias. A patterned magnesium layers were evaporated on biodegradable PLA
sheets by using a 3D printed mask. The circuits were assembled by heat sealing two magnesium patterned
sheets together to encapsulate the circuit structure. An inductive link is used to wirelessly detect the
resonance frequency of the circuit. The circuits were tested when immersed in de-ionised water and saline.
According to the tests, the designed resonator structure can be measured in aqueous environment. The
resonance of the tested circuit was observable at least for 51 hours. The concept still needs more
development to extend degradation time and to increase the stability during immersion.
1 INTRODUCTION
Implanting sensors into the body of a patient is for
fair reasons considered an extreme procedure
regardless of the benefits that can be achieved. The
main advance of the method is that the measurement
is not hindered by the sensor-skin interface. The
implanting procedure is more acceptable if it is done
at an unavoidable surgical operation due to injury or
disease. The threshold of using implantable sensors
is even lower if sensors are not permanent and thus
do not need a surgical removal operation. The
biodegradable sensors are developed to utilize this
niche.
One of the key features of an implantable sensor
is to access it by using a wireless link. RF or
inductive links are the most commonly used. The
structure of a passive resonance sensor that utilizes
an inductive link is very simple. Thus this method
has been utilized in many implantable (Collins 1967;
Chen et al. 2010) and biodegradable (Salpavaara et
al. 2012; Luo et al. 2014) sensor studies. A passive
resonance sensor is an LC circuit that is inductively
read by another coil. In this configuration, the
measurand is affecting the sensor coil or capacitor
thus altering the measured impedance of the reader
coil. Another approach is to link the measurand to
the losses in the LC circuit. To create a
biodegradable resonance circuit, coil and capacitor
structures are needed. Thus the methods for making
biodegradable patterned conductive and isolating
layers are needed. In addition, the sensor has to be
assembled and encapsulated without compromising
the made structures.
Conductive parts or layers can be made of
biodegradable metals (zinc, iron and magnesium) or
conductive polymers. The conductive polymers,
however, usually have too high resistivity for
passive resonant sensor applications. One
determining feature of the passive resonance circuit
is the Q-value. In order to achieve high Q-values the
resistance of the coil structure has to be small. For
this reason good conductivity is needed and the
fabrication method has to allow the preparation of
thick conductor layers. Bounty et al. made
conductive structures by using an electric discharge
machining to pattern 3 mm think iron and
magnesium sheets (Boutry et al. 2011). Luo et al.
have used an alternative method which utilizes
electroplating zinc and iron to create over 60 μm
thick coil wires (Luo et al. 2014). The evaporation
processes can be used to form patterned layers,
however, the time needed to deposit thick layers is a
problem. By using magnesium, high evaporation
rates can be achieved. Magnesium has been used to
form conductive areas to make wires, coils and
127
Salpavaara T., Ellä V., Kellomäki M. and Lekkala J..
Biodegradable Passive Resonance Sensor - Fabrication and Initial Testing.
DOI: 10.5220/0005255901270131
In Proceedings of the International Conference on Biomedical Electronics and Devices (BIODEVICES-2015), pages 127-131
ISBN: 978-989-758-071-0
Copyright
c
2015 SCITEPRESS (Science and Technology Publications, Lda.)
capacitors (Hwang et al. 2012).
The possible substrate materials for biodegra-
dable sensors can be polymers, silk and
bioresorbable glass. Biodegradable polymers are
divided into two groups: synthetic and those derived
from natural resources. There are also two
dominating degradation mechanisms: surface
erosion or bulk erosion. The surface eroding
polymers would make excellent candidates for
sensor fabrication because their water absorption
may be lower. One of the commonly used and well
tested group of biodegradable polymer polylactides,
(PLAs), unfortunately, degrades with bulk erosion.
Other suitable candidates for biodegradable polymer
substrates are polycaprolactone (PCL),
polyglycolide PGA, poly(3-hydroxybutyrate) (PHB)
and their co-polymers (Luo et al. 2014). Luo et al.
used Poly (L-lactide) PLLA as substrate material for
the passive resonance sensors. Another tested
substrate material for biodegradable sensors is
solution-casted silk (Hwang et al. 2012). They also
proposed the use of MgO
and SiO
2
as isolating
materials which are needed to coat conductors.
Besides individual fabrication processes, an
engineering problem arises with the assembly of the
fabricated structures. In comparison to a typical
silicon sensor fabrication, biodegradable structures
are prone to get compromised when new processing
steps are done over earlier layers. As an example,
magnesium layers break if any following steps
include water and many polymer structures will
deform if they get in contact with solvents. To solve
this problem, evaporation processes using physical
masks can be utilized (Hwang et al. 2012). This
method does not include photoresist masks which
have to be patterned and then removed by solvent.
Another method is to use embossing and lamination
techniques (Luo et al. 2014).
In this paper, ongoing research for methods of
designing and fabricating biodegradable passive
resonance sensors is discussed. The methods of
biodegradable sensor fabrication are developed and
combined in a novel way to create a biodegradable
resonator structure that can be modified to make
implantable sensors. The designed prototype
structure is assembled and tested in vitro.
2 FABRICATION METHODS AND
DESIGN
The principal idea of the presented biodegradable
resonator concept is to evaporate magnesium and
silicon dioxide on PLA sheets through a physical
mask. Then two sheets are stacked and combined to
a resonator structure by heat sealing. This method
encapsulates magnesium layers which are otherwise
prone to water. Another leading design feature is to
avoid galvanic through hole vias between the
conductive layers. Instead, conductive layers on the
separate sheets are connected by capacitors which
are formed when the sheets are heat sealed together.
The simplified cross-section of the design is shown
in Fig 1.
The electrical schematic diagram of the designed
circuit is presented in the Fig. 2a. This structure has
four coil and two capacitor structures. The schematic
does not include parasitic components. The
measured electrical behaviour of the presented
structure is similar to a simple RLC circuit and thus
can be roughly modelled as shown in Fig. 2b. The
resistance (Rs) is representing both the ohmic losses
in the magnesium layers and the dielectric losses
around and in the circuit. The inductor in the model
(Ls) represents the combination of four circular
planar coils (L1 to L4) on two opposing PLA sheets
and the inductance of the other connecting wires in
the circuit. In the final structure, four coils are
placed in a way that a changing uniform magnetic
field will induce currents that charge two main
capacitances in the circuit. The capacitor (Cs) in the
model is representing the capacitance of the two
electrode areas that are located inside the circular
coils and the combination of all parasitic
capacitances in the structure. The Fig. 2b also
illustrates the utilized wireless measurement
principle.
Figure 1: The cross-section of the resonator structure.
The idea behind this design can be derived from
a concept where a planar coil and a parallel-plate
capacitor are combined to an LC circuit by using a
through hole via between the layers. Instead of via,
another similar capacitor can be used to connect
layers. The capacitors are in series which has to be
taken account when the resonance characteristics are
designed. This kind of design, however, would be
unsymmetrical and in practice it would take almost
similar area as two planar coils and two parallel-
PLA
Mg
SiO2
50 nm
7.5 μ m
1 mm
BIODEVICES2015-InternationalConferenceonBiomedicalElectronicsandDevices
128
plate capacitor electrodes per layer. Thus the set of
four coils and two capacitors is adopted in this
design to maximise the inductance in the circuit.
Figure 2: The electrical configuration of the fabricated
resonance circuit (a). The measurement setup and the
simplified model of the resonance circuit (b).
The substrate PLA sheets are made of PLA
(Purac Corbion Purasorb PLD962). First, granules
are melt processed to a 10 mm diameter rod. Then
the rod is cut to smaller pieces which are
compressed to 500 μm thick sheets by using elevated
temperature and simultaneous pressure. One side of
the polymer is compressed against glass sheet to
make the surface of the polymer smooth enough for
the evaporation process. Finally, the sheets are cut to
40 mm by 30 mm pieces.
The magnesium is patterned by using physical
masks during the evaporation process. In this case,
the masks were created by extrusion type 3D
printing. This decision was made to test the
capabilities of 3D printing and to accelerate the
prototyping process. The extrusion path was
manually designed to be as continuous as possible
since the most of the irregularity of the 3D printed
object occurs when the extrusion is started or
stopped. Also one design criterion of the designed
masks is that the objects have to support themselves.
In practice this means that printed mask has to be
continuous and there cannot be parts that sag when it
hung on the roof of the evaporation chamber. The
99.99% pure magnesium was evaporated on the
PLA sheets as a 7.5 μm ± 0.3 μm thick layer. The
mask and the evaporated coil pattern are shown in
Fig. 3. The magnesium layer was coated with 50 nm
SiO
2
-layer which acts as an electrical isolator
between the capacitor plates.
Figure 3: The 3D printed mask and corresponding
magnesium pattern on PLA sheet.
Figure 4: The magnesium patterned PLA sheet (a) and
resonator structure assembly (b).
A resonator structure was assembled by stacking
two patterned PLA sheets (Fig. 4) together with
patterned sides facing each other. In addition, two
patterned layers are placed in a way that coils on the
opposite sheets turn opposite directions. The sheets
were pressed together in a workbench and each edge
was trimmed and heat sealed to waterproof the
structure.
3 RESULTS
The impedance magnitude and phase of the
resonators were measured in air by using an
impedance analyser (Agilent 4396B with impedance
L1
Mutual
inductance
Lr Ls
Measurement
Cs
Rs
L2
L3 L4
C1 C2
a
b
BiodegradablePassiveResonanceSensor-FabricationandInitialTesting
129
test adapter) and an external coil (diameter 40 mm).
The impedance of the reader coil was also measured
separately and this baseline was removed from the
measurements. The two sample sensors have
similarly shaped responses curves at clearly separate
frequencies (Fig. 5). The frequencies of the phase
dips (f
p
) of these sensors are at 50.8 MHz and 58.9
MHz.
Figure 5: The magnitude and phase responses of the
resonance sensors measured in air.
The sensor 1 was immersed in 30 ml of 9 mg /
ml saline and sensor 2 in 30 ml of de-ionized water.
The sensors were supported from the edges in a way
that there was a 1 mm thick liquid layer under and
over the tested object. The temperature of the test
environment was 22 Cº. The changes of f
p
are shown
in Fig. 6. The initial drop due to the immersion is
fairly similar in both cases. This is followed by the
expected decrease of the f
p
. After two hours of
immersion, the f
p
starts to increase again. This
continues until roughly 6
th
hour in the case of sample
one. After 6 hours, signals start to degrease again
and in the case of sample 2, the decrease of f
p
was
faster. The measurement of the sample 2 was
stopped after 8
th
hour. The f
p
of sensor 1 increased
dramatically between ten and twenty hours. The
resonance was still clearly detectable until 51
st
hour.
Figure 6: The frequencies of the phase dips of the tested
sensors during immersion to de-ionized water and saline.
The degradation was also visually monitored (Fig.
7). There are no visual signs of degradation during
the first 24 hours. The magnesium conductors start
to show degradation after 48 hours of immersion to
saline. After one week immersion in saline, the
corrosion is clearly visible all over the magnesium
patterns.
Figure 7: The degradation of the magnesium conductors
begins to be visually detectable after 2 days immersion to
saline.
4 DISCUSSIONS
The proposed structure for a biodegradable resonator
has been successfully demonstrated. The scheme to
avoid galvanic through hole vias can be realized.
This simplifies the manufacturing process and the
presented fabrication method has only a few process
steps. The optimization of the presented structure
may yield better results; however, modelling all the
parasitic components in the structure is a
complicated task.
The 3D printed physical masks worked well in
this application. The performance of the printers will
improve in coming years which may encourage
using them for mask fabrication. However, it should
be noticed that masks with fine details were only
used to pattern magnesium in this study and their
temperature limitations should be considered if used
in other applications.
The general design is not especially sensitive to
fabrication tolerances since the two made resonators
have similar response curves. The difference in the
resonance frequencies in air is not too significant
considering manual assembly and the 3D printed
masks. The main reason for the difference in the
20 30 40 50 60 70 80
-5
0
5
Frequency (MHz)
Magniture (ohm)
20 30 40 50 60 70 80
-15
-10
-5
0
Frequency (MHz)
Phase (deg)
Sensor 1
Sensor 2
-10 0 10 20 30 40 50 60
-5
0
5
10
15
20
25
30
Time (h)
Frequency (MHz)
Saline
DI-water
BIODEVICES2015-InternationalConferenceonBiomedicalElectronicsandDevices
130
resonance frequencies of this design is estimated to
be the parallel plate capacitors assembly. In addition
to alignment errors, the structure is sensitive to
assembly pressure and the mechanical properties of
the substrate PLA sheets. As process steps are
developed, the resonance frequencies can set more
accurately and variation will be smaller.
The main drawbacks of sample prototypes are
the short durability in aqueous environment and the
instability of the resonance characteristics due to the
water absorption of PLA. The changes in the
resonance characteristics of the tested circuits were
so large that the measurements have to be verified
with more parallel samples before any better
conclusions can be drawn. Also the hydrolysis test
of the device has to be longer and samples have to
be measured with more dense intervals. It can be
estimated that instability is caused by the absorbed
water which changes the electrical fields inside the
sensor structure. Other likely cause for instability is
that unfixed capacitors plates will move as PLA
absorbs water and swells. This may explain the large
increase in the resonance frequency after first 10
hours of immersion.
The current study proved that the PLA sheets
alone do not provide adequate encapsulation for
magnesium. As a quick fix, magnesium oxide can be
evaporated under and top of the magnesium layer
since this material has been used to modify the
degradation (Hwang et al. 2012). The magnesium
can also be replaced with more slowly degrading
metal like zinc, this, however, is unlikely to fix the
instability problem and zinc has smaller conductivity
compared with magnesium. Replacing PLA with a
surface eroding polymer can be a valid option since
the designed structure worked well in test conditions
before water had time to absorb in the structure.
Another option is to modify the water absorption of
the PLA for example with a coating.
The structure is sensitive to pressure variations
due to the mechanical changes in the assembled
capacitors. This can be utilized in order to make
pressure sensors or the sensitivity may be mitigated
by filling the empty space that is left in the structure.
Replacing the air in the structure with biodegradable
hydrophobic material and fixing the capacitors
plates together will make the structure more durable
and may improve stability during immersion. In
present design, only the edges of the PLA sheets
were fixed.
The research will continue by modifying the
designed resonator structure. The short-range goal is
to develop a resonator that is stable for months.
Then the design will be modified to be sensitive to a
specific measurands like pressure or permittivity
outside of circuit.
REFERENCES
Boutry, C.M., Chandrahalim, H. & Hierold, C., 2011.
Characterization of RF Resonators Made of
Biodegradable Materials for Biosensing Applications.
Procedia Engineering, 25, pp.1529–1532.
Chen, P.-J. et al., 2010. Wireless Intraocular Pressure
Sensing Using Microfabricated Minimally Invasive
Flexible-Coiled LC Sensor Implant. Journal of
Microelectromechanical Systems, 19(4), pp.721–734.
Collins, C.C., 1967. Miniature passive pressure transensor
for implanting in the eye. IEEE transactions on bio-
medical engineering, 14(2), pp.74–83.
Hwang, S.-W. et al., 2012. A physically transient form of
silicon electronics. Science (New York, N.Y.),
337(6102), pp.1640–4.
Luo, M. et al., 2014. A Microfabricated Wireless RF
Pressure Sensor Made Completely of Biodegradable
Materials. Journal of Microelectromechanical
Systems, 23(1), pp.4–13.
Salpavaara, T. et al., 2012. Biodegradable encapsulation
for inductively measured resonance circuit. In
Bioinformatics & Bioengineering (BIBE), 2012 IEEE
12th International Conference on. Larnaca: IEEE, pp.
11–13.
BiodegradablePassiveResonanceSensor-FabricationandInitialTesting
131