product of the diffusion coefficient and solubility
coefficient. The units of Dk value are 10-11 (cm3 O2
cm)/ (cm3 sec mmHg) or "barrier". The higher is the
value of Dk, the better is the oxygen permeability of
CL. In addition, the Oxygen transmissibility (Dk/t) of
the local region of CL is expressed as oxygen
permeability (Dk) of the material of CL is divided by
the thickness (t) of CL.
There are two commonly methods to measure the
Dk or Dk/t of CL, the polarographic and coulometric
technologies (Fatt and Chaston, 1982, Refojo, et al.,
1977, Refojo and Leong, 1979, Brennan, et al., 1986,
Compañ, et al., 1996, Paterson and Doran, 1986 and
González-Méijome, et al., 2008). The polarographic
technique has often been used to determinate the Dk
coefficient of hydrogel CL, which placed directly on
the Clark-type electrode, based on the oxygen flux
through the CL. Practically, this technique has its
limitations during measuring Dk, such as the so-
called boundary-layer effect and edge effect in the
polarographic method. The boundary-layer effect
leads to the underestimation the Dk/t of sample due
to the difference of oxygen partial pressure on the two
side of sample. In addition, the edge effect means that
the lateral diffusion of oxygen occurs if the test area
is not the same on both sides of sample and results in
the Dk value of sample under test is slightly higher
than the real value. Despite these shortcomings in the
measurement of polarographic method, but it can use
some experimental procedure to amend its measured
value to close the actual value. For example, the
measuring values could be corrected by measuring
the samples of different thickness for the boundary-
layer effect (Compañ, 2002), or be multiplied by a
proper correction factor for the edge effect. In the
coulometric method, a nitrogen carrier gas flows
around the lens and transports the permeated oxygen
to the oxygen detector which produces an electrical
current, wherein the magnitude of oxygen through the
film is proportional to the amount of oxygen. The
oxygen gas transmits from upper chamber through
CL film into lower chamber during the permeability
process. However, the sample has a dehydration
effect during test when it exposed to air; therefore, the
coulometric method is not applied to the Dk
measurement of hydrogel CL because the water
content can cause changes in oxygen permeability.
The above mentioned two methods are both well
defined in the ISO standards (ISO 9913-1, 1996 and
ISO 9913-2, 2000), which also referred to the
restrictions on the use of two methods. The
polarographic method can only measure the CLs of
less than 100 Barrier, and the coulometric method
cannot be applied to the hydrogel CLs. However, the
polarographic (Fatt) and coulometric method in ISO
can be adopted to determinate the oxygen permeation
through all types of contact lenses, except the high Dk
silicone based CLs. Therefore, these standards 9913-
1&2 have now been withdrawn and replaced by (ISO
18369-3, 2006). The ISO 18369-3 specifies the
methods of testing the physicochemical properties of
CLs, which are extraction, rigid lens flexure and
breakage, oxygen permeability, refractive index and
water content. Therein, the soft CLs can also be made
of non-hydrogel materials, such as the silicone
elastomers. Based on the high performance liquid
chromatography (HPLC-EC), (Oberndorf and
Wilhelm, 2003) uses the reductive electrochemical
detection to determine oxygen at nanomolar levels,
the method had not only lens dehydration, but can
minimize the edge and boundary layer effects, the Dk
values of rigid and soft CL can be determined in the
same manner with good reproducibility. The above
mentioned measuring methods for oxygen
permeability of CLs are based on the electrochemical
or vacuum infiltration model, so the measuring results
are often not consistent due to the different measuring
method or instruments.
Therefore, this study presents a non-contact
method that can reduce the measurement error came
from the environment or instrument layout, and to
evaluate the WC and Dk/t of CL in different manner
from the absorption spectrum of CL. Therein, the full
wavelength of light was provided from the halogen
lamp, and the variation of light spectrum was
measured from the spectrometer that facing to the
light source. In addition, five types of CL with
different WC and Dk/t were employed to observe the
variation of intensity of transmission light in 8 minute
after wearing to the experimental setup.
2 FUNDEMENTAL THEORY AND
EXPERIMENT SYSTEM
2.1 Relationship between Water
Content and Oxygen Permeability
of Contact Lens
In general, the Dk/t of CLs is positive correlation to
the WC of CLs, (Hadassah and Sehgal, 2006)
presented a measurement that allowing the oxygen to
pass through the lens material and investigate the
oxygen permeability and transmissibility of contact
lenses of different thickness and curvature. Therein,
the expelled oxygen gas was collected by the
dissolution in ethanol and measured by the titration of