Image Quality Comparison between Synthetic 2D Mammograms Obtained with 15º and 40º X-ray Tube Angular Range: A Quantitative Phantom Study
R. M. Tucciariello, R. M. Tucciariello, P. Barca, P. Barca, M. E. Fantacci, M. G. Bisogni, D. Caramella, V. Rosso, V. Rosso, R. M. Tucciariello, R. M. Tucciariello, A. C. Traino, M. E. Fantacci, M. G. Bisogni
2020
Abstract
In this work we present an image quality comparison between synthesized mammograms (SMs) obtained from Digital Breast Tomosynthesis (DBT) acquisitions with 15° (SM15) and 40° (SM40) X-ray tube angular range. In fact, since wide-angle DBT is characterized by a better spatial resolution in depth but also by worse performance in detecting microcalcifications than narrow-angle DBT, an objective image quality analysis of SM images could be of pratical interest. Four phantoms were employed in this study and their images were acquired using an Amulet Innovality mammographic device. The image quality comparison was conducted by evaluating spatial resolution, contrast and noise properties of the images. Our results show that SM40 images are characterized by better spatial resolution performance than SM15 in terms of Modulation Transfer Function but also by worse performance in the detection of low-contrast details. In fact, higher contrast-to-noise ratio values were obtained with SM15 than with SM40. Noise properties of the images were also investigated through the Noise Power Spectrum (NPS) calculation: no differences in NPS shapes were found in both modalities, while noise magnitude results significantly different. In addition, Signal-to-Noise Ratio (SNR) spatial distribution evaluation was assessed by computing SNR maps, in which different pattern were observed.
DownloadPaper Citation
in Harvard Style
Lamastra R., Barca P., Bisogni M., Bisogni M., Caramella D., Rosso V., Tucciariello R., Traino A., Fantacci M. and Fantacci M. (2020). Image Quality Comparison between Synthetic 2D Mammograms Obtained with 15º and 40º X-ray Tube Angular Range: A Quantitative Phantom Study. In Proceedings of the 13th International Joint Conference on Biomedical Engineering Systems and Technologies (BIOSTEC 2020) - Volume 2: BIOIMAGING; ISBN 978-989-758-398-8, SciTePress, pages 184-191. DOI: 10.5220/0009147601840191
in Bibtex Style
@conference{bioimaging20,
author={R. M. Lamastra and P. Barca and M. G. Bisogni and M. E. Bisogni and D. Caramella and V. Rosso and R. M. Tucciariello and A. C. Traino and M. G. Fantacci and M. E. Fantacci},
title={Image Quality Comparison between Synthetic 2D Mammograms Obtained with 15º and 40º X-ray Tube Angular Range: A Quantitative Phantom Study},
booktitle={Proceedings of the 13th International Joint Conference on Biomedical Engineering Systems and Technologies (BIOSTEC 2020) - Volume 2: BIOIMAGING},
year={2020},
pages={184-191},
publisher={SciTePress},
organization={INSTICC},
doi={10.5220/0009147601840191},
isbn={978-989-758-398-8},
}
in EndNote Style
TY - CONF
JO - Proceedings of the 13th International Joint Conference on Biomedical Engineering Systems and Technologies (BIOSTEC 2020) - Volume 2: BIOIMAGING
TI - Image Quality Comparison between Synthetic 2D Mammograms Obtained with 15º and 40º X-ray Tube Angular Range: A Quantitative Phantom Study
SN - 978-989-758-398-8
AU - Lamastra R.
AU - Barca P.
AU - Bisogni M.
AU - Bisogni M.
AU - Caramella D.
AU - Rosso V.
AU - Tucciariello R.
AU - Traino A.
AU - Fantacci M.
AU - Fantacci M.
PY - 2020
SP - 184
EP - 191
DO - 10.5220/0009147601840191
PB - SciTePress