Analysis of Mechanical Properties of Dialysis Sets
Klara Fiedorova
1
, Martin Augustynek
1
, Frantisek Fojtik
2
and Milada Hlavackova
2
1
VSB Technical University of Ostrava, Faculty of Mechanical Engineering,
Department of Cybernetics and Biomedical Engineering, Ostrava, Czech Republic
2
VSB Technical University of Ostrava, Faculty of Mechanical Engineering,
Department of Applied Mechanics, Ostrava, Czech Republic
Keywords: Blood Dialysis Set, Haemodialysis, Mechanical Properties of PVC, Mechanical Tests.
Abstract: Dialysis sets are part of the dialysis circuit and are used every day. It is important to find out how the
dialysis sets behave under load and whether they retain their typical properties, which are the strength and
flexibility. Therefore, the mechanical properties of the dialysis sets are evaluated depending on the selected
physical parameters. For this reason, the sets are subject to temperature changes and changes in the pressure
of the peristaltic pump occlusion cylinders. To evaluate the mechanical properties of the dialysis sets, the
sets were subjected to stress tests and subsequent tests. The results were processed and analysed.
1 INTRODUCTION
Work deals with dialysis sets. Set are used during
blood dialysis. They ensure the transfer of blood
flowing from the patient's body to a blood-purifying
dialysis machine and to return the purified blood
back to the patient's body. The material from which
the dialysis sets are made is very important. Material
biocompatibility is an important component for the
success of hemodialysis. The set is not to produce
thrombogenic, inflammatory, immune or toxic
reactions. The dialysis sets must meet the criteria,
such as mechanical strength of material and optical
properties. Therefore it made of polyvinyl chloride.
Due to their chemical composition today they are
condemned. (Augustynek et al., 2014), (Neergaard
et al., 1971), (Liao et al., 2003) Therefore, the
products emerging from alternative materials. But
the use of PVC medical devices is still very
important, because some of the alternative materials
do not have the necessary properties. (Kostic et al.,
2017)
The result of the work should be an analysis of
the mechanical properties of dialysis sets. These
properties are dependent on physical parameters.
Mechanical properties mean the behaviour of
materials under the influence of external mechanical
forces. Sets are exposed to deformation by pressure
of occlusion cylinders and change in temperature of
saline. Dialysis sets must be strength, elasticity and
malleability to withstand stress tests. It is assumed
that under optimal conditions the dialysis sets meet
the standards. However, if the set goes through a
series of loads, measurements should demonstrate
the change of mechanical properties of dialysis sets.
This work focus on a group of AV dialysis sets,
which are most commonly used today. Because
dialysis sets are used during blood dialysis, their
properties may affect the quality of therapy.
Therefore, in this work, the sets undergo a load and
subsequent mechanical tests of physical properties.
By analyzing the obtained data we are able to
determine the limit of the mechanical load so as to
avoid a malfunction. (Brugger, 1999), (Beisser and
Oesterreich, 2018)
1.1 Properties of Dialysis Sets
Set should guarantee a safe path of blood from the
patient's body and return cleaned blood back. It
belongs to special medical supplies, which must
meet certain characteristics. In order to be able to
fulfil its function, set made from polyvinyl chloride.
To ensure the necessary features of the sets, PVC is
softened by the plasticizer during production. The
most commonly used plasticizer is di-ethylhexyl
phthalate (DEHP hereinafter). (Neergaard et al.,
1971), (Kostic et al., 2017)
Another reason for using PVC in making dialysis
sets is transparency and clarity. During the actual
228
Fiedorova, K., Augustynek, M., Fojtik, F. and Hlavackova, M.
Analysis of Mechanical Properties of Dialysis Sets.
DOI: 10.5220/0007555302280234
In Proceedings of the 12th International Joint Conference on Biomedical Engineering Systems and Technologies (BIOSTEC 2019), pages 228-234
ISBN: 978-989-758-353-7
Copyright
c
2019 by SCITEPRESS Science and Technology Publications, Lda. All rights reserved
blood dialysis is very important to check it by eye,
the nurse examines the blood. Thanks to these
characteristics nurse can see harm of dialysis tubing
or may capture a blood clot in the blood. (Di Landro
et al., 2005), (Latini et al., 2010)
Dialysis sets are important parts of the dialysis
device. There are two pressures on the dialysis set
during blood dialysis. The external pressure is
caused by the peristalsis pump. The effect of the
occlusion cylinders is to inject blood into the device
and further drive the circuit. Thus, a flow of a
certain value is created. Blood flowing through the
dialysis set produces internal pressure. During the
measurement, the temperature of the solution is
warmed to the temperature of the blood, ie it has a
temperature of 37. Therefore, the effect of the
temperature on the properties of the sets must be
considered. Optimum physical parameters under
which the dialysis is carried out, the temperature of
the solution through 37 and speed peristaltic pump
300 ml/min. (Rose et al., 2012), (Ronco and Levin,
2004), (Cihak and Augustynek, 2013), (Jeppsson
and Losell, 1994), (Brugger et al., 1997)
1.2 The Important Mechanical
Properties
The mechanical properties are evaluated in the work.
During testing, the sets undergo stress. The dialysis
set has to be characterized by basic features such as
strength, elasticity, toughness, malleability. We use
these properties in real use fatigue resistance.
Temperature plays a role in the material properties;
the influence of certain temperatures may cause a
change in the material structure. (Pelleg, 2013),
(Dowling, 2007)
The dialysis sets are produced from plasticized
polyvinyl chloride. Due to this fact, they are
characterized by strength, flexibility, elasticity, etc.
Flexibility - elastic deformation before the
breach, evaluated using the modulus of
elasticity and stress.
Strength - material resistance to deformation
and breach of external forces.
Plasticity - the material retains permanent
deformation induced by external forces,
assessed using the relative size of the
permanent deformation prior to the breach.
Toughness - resistant to high tension, this
feature is dependent on the strength and
plasticity. The measure of the work required
to deform a violation.
(Pelleg, 2013), (Dowling, 2007)
1.3 Standards Applied
The standard EN ISO 8638 Cardiovascular implants
and extracorporeal systems - Extracorporeal blood
circuit for haemodialyzer, haemodiafilter and
haemofilter deals with dialysis sets. The standard
lists the requirements for dialysis sets and tests that
test the dialysis sets. Tests are performed on the
testing machine, which will assess the ability of a
material to withstand the required stress. It records
the force which must be exerted in order to break the
material. Other values are modulus and elongation
of the material. (ČS EN ISO 6838, 2014)
1.4 Testing of Mechanical Properties
According to the standard above, dialysis sets are
subjected to a tensile test. This test is part of
mechanical static tests. The tensile strength is
performed on all materials because, thanks to the
measured values, we can further count the structural
elements and choose the appropriate material. The
tensile strength test indicates the force exerted,
variable elongation, elongation and material
constriction.
2 TESTING OF DIALYSIS SETS
Measurements were measured in were school lab
because the laboratory has a dialysis machine
Fresenius 4008 and other essential components that
are, dialysis sets Basic Line AV, dialyzers AV600S
and dialysate. Dialysis apparatus with dialyzer and
bag can be seen in Figure 1.
Figure 1: A – Bibag, B – Hemodialysis Fresenius 4008, C
– Dialysator AV 600S (Cihak and Augustynek, 2013).
Measurements are divided into several series. The
series differs in the used and set parameters. The
combination and the values of these parameters are
given in the following section. After the part of the
measurement has been completed, tensile tests are
carried out. The obtained data determine the
mechanical properties of the dialysis kits. The
assumed, that dialysis sets after a series of
Analysis of Mechanical Properties of Dialysis Sets
229
measurements when exposed to the highest pressure
and the highest temperature should have the greatest
degree of flexibility. Thus, there should be a greater
extension and the equation:

∙
∙
(1)
where is force, is the modulus of tensile
elasticity,
is tensile strength, is original length,
is elongation length. We can deduce that it will
increase the maximum force expended and the
modulus of elasticity decreases as the module
increases with stretching flexibility. (Pelleg, 2013),
(Beisser and Oesterreich, 2018), (Rose et al., 2012)
2.1 Determination of Possible Influence
of Changes of Physical Parameters
on Mechanical Properties of
Dialysis Sets
The tensile test is performed during the production
of the set. It is assumed that the sets will be used
under standard conditions. Such a set must have
certain features, such as flexibility, tightness,
plasticity, according to the standard. Therefore, the
aim of this work is to verify the effect of certain
physical parameters on the properties of the set.
(Hochman et al., 2015)
Work focuses on the influence of temperature
and pressure. The effect of temperature on the
mechanical properties of the set is derived from its
structure. PVC is made up of linear or slightly
branched chains. It is assumed that by varying the
temperature, the internal structure of the set will
expand or retract. At higher temperatures it will be
more flexible. On the contrary, the strength of the set
will fade. The effect of pressure on the mechanical
properties of the set will only affect the place of
direct action, not the length of the whole set. If the
pressure is increased, the material will weaken at the
point of application. This reduces the strength.
(Hochman et al., 2015)
2.2 Workflow Proposal
Since we cannot work with the biological material in
the laboratory, it is measured by with the
physiological solution heated to the temperature of
human blood, which is 37 . Instead of connecting
to the vascular access use of circular bags, which
join dialysis sets. The saline which flows through the
circuit, it will drain back into the bag. The important
part is ripper for determining the mechanical
properties of dialysis sets, which we will evaluate.
Measurements must be carried out under steady
conditions (pressure, temperature, humidity).
Measurements on one set is always carried out only
once because dialysis sets are used for single use. In
the experimental measurement, two measuring
parameters are changed, namely the temperature of
the saline solution and the speed of the peristaltic
pump, thereby achieving a change in pressure.
Temperatures are used: 36.5ºC, 37ºC, and 37.5ºC
and speed of the peristaltic pump: 200 ml/min,
300 ml/min and 450 ml/min. Each combination of
measurement conditions shall be subjected to a four-
hour measurement.
Figure 2 shows a dialysis device. Part of a
dialysis machine belonging to the arterial part is
marked by the letters B, C and D. Part of dialysis
device marked with B is a peristaltic blood pump,
this section includes the insulin pump segment.
Under C is located the arterial blood pressure sensor
to which the set up via the transmitter. The heparin
pump is under D. The arterial part of the set also
includes the arterial chamber is located under the
letter E. The letters G, H, CH are the venous part of
the set. Point G shows the air sensor, here is placed
venous chamber. Below the H, there is a fuse in the
circuit that interrupts the blood flow in case of an
alarm. The pressure sensor can be found under the
letter CH. Letter A indicates the monitor. Under the
letter I, a physiological saline bag is found. Letter F
denotes a part, to which the inlet and outlet hoses are
attached, they are stored under the letter J. The
dialysis solution is made up of chemicals, which are
stored in parts marked K, L.
Figure 2: Dialysis machine.
2.3 Mechanical Test
Mechanical tests were carried out at VŠB-TUO at
the Faculty of Mechanical Engineering. It was tested
BIODEVICES 2019 - 12th International Conference on Biomedical Electronics and Devices
230
on the device Testometric M500-50CT (see Figure
3). Those parts that were exposed to stress were
tested. In the case of experimental measurements, it
is peristaltic pump segment. Sets is characterized by
an inner diameter of 8 mm and an outside diameter
of 12 mm measured. The length of the test sample
was 6 cm. The sample was expanded at
100 mm/min. The samples were attached to the vice
of the instrument, each side being held at 0.5 cm.
The whole machine is controlled by software, where
the computer sets up the appropriate parameters
(length of the sample, the outer and inner diameter,
speed). After setting the appropriate parameters, the
machine starts.
After running a device the test material is
tensioned. The material to be tested should break
with a certain extension. The whole course of testing
is recorded in a graph of force versus extension.
Figure 3: Testing of dialysis kits.
3 DATA PROCESSING
Data obtained from material testing on mechanical
tests, calculations and measurements will be
processed and evaluated using tables and graphs.
Prolongation will depend on selected physical
quantities (flow and temperature).
The following variables (shows in Table 1) are
used in the work.
Table 1: Overview of the variables.
Quantity: Unit:
Temperature of saline solution
[]
Speed peristaltic pump blood [ml/min]
Mechanical tests are extension [mm]
Maximum developed force [N]
Young's modulus [MPa]
Length is expressed [mm]
3.1 Dependence of Extension on
Temperature
The chart (Figure 4) shows the dependence of the
extension on the saline temperature. Three
temperatures were set for each used speed. On the X
axis the values of the saline temperature [] were
plotted, and an extension [mm] was recorded on the
Y axis. When the peristaltic pump speed was set at
200 ml/min, the largest extension from the 60 mm
sample was 367.071 mm, this sample was recorded
at 37℃. For 300 ml/min was the largest extension
414,57 mm (37℃) and for 450 ml/min it was
350,573 mm. We can see that prolonging varies with
temperature, which satisfies the assumption. It is
apparent from the graph that the least elongation has
200 ml/min samples and gradually increase with an
increasing flow. At a speed of pump 450 ml/min, the
extension does not change much. In this graph, we
can see that at the temperature of 37 °C, the test
sample is most prolonged.
Figure 4: The dependence of prolongation on temperature.
3.2 Dependence of Maximum Force on
Temperature
The graph (Figure 5) shows the dependence of
maximum force on the temperature of the saline.
Values were obtained by the same procedure as in
the previous chapter. On X axis, the saline
temperature [°C] is showed, and the maximum force
[N] is recorded at Y axis. From the values in the
graph it is evident that as the temperature of the
solution increased, the force that was needed to
break the test sample decreased. In the sample
obtained from the measurement at 200 ml/min, the
smallest applied force was 407.9 N at 37.5°C. For
300 ml/min is the smallest applied force 458.9 N at
36.5°C and for 450 ml/min is force 458,3 N at
37.5°C. The force applied to the test sample was the
smallest in the sample test at a pump speed of
200 ml / min. However, the maximum force applied
to the samples at a speed of the pump 450 ml/min is
less than the maximum force that was applied to the
samples measured at 300 ml/min. It can be seen in
this figure (Figure 5) that the samples measured at
37 ° C are the most stable.
Analysis of Mechanical Properties of Dialysis Sets
231
Figure 5: The dependence of the maximum forces on the
temperature.
3.3 Dependence of Modulus on
Temperature
The graph (Figure 6) shows the dependence of the
modulus on the temperature. The temperature of the
saline [°C] was plotted on the X axis and a modulus
of elasticity [MPa] was recorded on the Y axis.
Considering that the modulus of elasticity is the
greater the stronger the test material, in theory, in
our case, this modulus of elasticity should decrease.
Because sets that have been exposed to more stress
become more flexible. For samples generated at
200 ml/min, the maximum modulus of elasticity is
91.166 MPa at 37°C. For speed 300 ml/min, the
highest elasticity model was 54.7 MPa at 37°C. The
highest modulus for 450 ml/min is 53.773 MPa at
36.5°C. This chart (Figure 6) we confirm
assumption, namely the gradual decrease of the
modulus of elasticity with the increasing stress of
the set. Only the values listed on the sets measured
at 36.5°C, are different.
Flow rate 200 ml/min Flow rate 300 ml/min Flow rate 450 ml/min
Figure 6: The dependence of elastic modulus on
temperature.
3.4 Dependence of Extension on Flow
Rate
The graph (Figure 7) shows the dependence of the
elongation on the speed of the peristaltic pump. At
each temperature, each measured speed was used.
Flow rates [ml/min] were plotted on the X axis, the
elongation [mm] was plotted on the Y axis. The
increasing flow rate should result in prolongation of
the test specimens. Because the material becomes
more flexible. We can see, that for samples
measured at 36.5°C, the longest extension is
368.106 mm at 450 ml/min. The largest extension
for samples with a temperature of 37°C is
414.579 mm at a pump speed of 300 ml/min. The
largest extension for samples with a temperature of
37.5°C is 339.464 mm at a pump speed of
450ml/min. The graph shows that the sets exposed
to 37 ° C are out of the assumption.
Temperature 36,5 °C Temperature 37 °C Temperature 37,5 °C
Figure 7: Dependence of elongation on flow rate.
3.5 Dependence of the Maximum Force
on Flow Rate
The graph (Figure 8) shows the dependence of the
maximum applied force on the flow rate. The flow
velocity values [ml/min] were plotted on the X axis,
and a maximum force [N] was recorded on the Y
axis. Higher blood pump speeds should result in a
weakening of the material. We can see, that for
samples measured at 36.5°C, the maximum force is
479.6 N at 450 ml/min. The maximum force for
samples with a temperature of 37°C is 501.1 N at a
pump speed of 300 ml/min. The maximum force for
samples with a temperature of 37.5°C is 483.2 N at a
pump speed of 300 ml/min. We can see that the
effect of flow does not have such an effect on the
maximum force used, because the measured values
are quite similar. The data of samples measured at
37°C and 300 ml/min are again different.
BIODEVICES 2019 - 12th International Conference on Biomedical Electronics and Devices
232
Temperature 36,5 °C Temperature 37 °C Temperature 37,5 °C
Figure 8: The dependence of the maximum forces on the
flow rate.
3.6 Dependence of Modulus on Flow
Rate
In the graph (Figure 9) we can see the evaluation of
the dependence of the modulus on the flow rate. The
flow velocity values [ml/min] were plotted on the X
axis and the modulus of elasticity [MPa] on the Y
axis. For samples generated at 36.5°C, the maximum
modulus of elasticity is 53.773 MPa at 450 ml/min.
Samples measured at 37°C and 37.5°C using a
different flow rate meet the theory that the elastic
modulus decreases with increasing effort. The values
of all measured samples shows (Figure 9) that the
values for the samples exposed at 37°C, again stands
out. We can see that the model flexibility is
diminished with increasing temperature.
Figure 9: Dependence of the elastic modulus on the flow
rate.
4 DISCUSSION
All combinations of temperatures and flows were
tested, one by one. A part of the dialysis sets that
was exposed to peristaltic pumps was tested.
Considering the assumptions set, the samples
obtained at temperature 37°C do not meet the
requirements. But the remaining samples confirm
the assumptions. These fluctuations can be justified
by the fact that hemodialysis is normally used at
37°C and this temperature influences the production
of sets.
The elongation of the sample and the applied
force applied to the specimen increases with the
load, ie the flow temperature and the flow rate. In
addition, the elastic modulus of test samples which
were subjected to increased strain gradually
decreases. Expansion between 200 - 400 mm was
recorded, with increasing expansion increasing the
maximum applied force, which was measured in the
range of 400 - 500 N. The range of 40 to 100 MPa
include modulus of elasticity.
Said modulus could not be verified by
calculation, as Hook's law applies only to a small
linear part that cannot be deducted from the chart.
Beyond this, there is a large deformation of the
material.
5 CONCLUSION
From the analysis of the experimental measurements
it is possible to confirm the assumptions that the
change of the physical parameters (temperature of
the physiological solution and the flow) affects the
properties of the dialysis sets. Changing the
mechanical properties of dialysis sets is not so
significant as to affect the safety of the use of sets.
The sets proved to be a very solid and durable
material on mechanical tests. Situations that may
occur in health care facilities under the supervision
of professional staff will not expose a set of
conditions to fatal failure of the material.
ACKNOWLEDGEMENTS
The work and the contributions were supported by
the project SV4508811/2101Biomedical
Engineering Systems XIV'. This study was also
supported by the research project The Czech Science
Foundation (GACR) 2017 No. 17- 03037S
Investment evaluation of medical device
development run at the Faculty of Informatics and
Management, University of Hradec Kralove, Czech
Republic. This study was supported by the research
project The Czech Science Foundation (TACR)
ETA No. TL01000302 Medical devices
development as an effective investment for public
and private entities.
Analysis of Mechanical Properties of Dialysis Sets
233
REFERENCES
R. J. Rose, M.J. Priston, A.E. RigbyJones, J.R. Sneyd,
2012. The effect of temperature on di (2ethylhexyl)
phthalate leaching from PVC infusion sets exposed to
lipid emulsions. In Anaesthesia, 67(5), 514-520.
Z. Liao, C.K. Poh, Z. Huang, P.A. Hardy, W.R. Clark, D.
Gao, 2003. A numerical and experimental study of
mass transfer in the artificial kidney. In Journal of
biomechanical engineering, 125(4), 472-480.
C. Ronco, N.W. Levin, 2004. Hemodialysis, Vascular
Access, and Peritoneal Dialysis Access. In Karger
Medical and Scientific Publishers, (Vol. 142).
J. Cihak, M. Augustynek, 2013. Infuzní technika a
hemodialyzační technika a technologie. In Ostrava:
VŠB - Technická Univerzita Ostrava, p. 373, 2013,
ISBN 978-80-248-3100-8.
J. Pelleg, 2013. Mechanical properties of materials. In
Springer, c2013. Solid mechanics and its applications,
ISBN 978-94-007-4341-0.
N.E. Dowling, 2007. Mechanical behavior of materials:
engineering methods for deformation, fracture, and
fatigue. In Upper Saddle River: Prentice Hall, 3rd ed.
2007. ISBN 0-13-186312-6.
ČSN EN ISO 8638 Kardiovaskulární implantáty a
mimotělní systémy: Mimotělní krevní okruhy pro
hemodialyzátory, hemodiafiltry a mofiltry, 2014.
V. Hochman, J. Cihak, M. Augustynek, 2015. Interaction
of Infusion Set and Volumetric Infusion Pump and
Their Impact on the Quality of Treatment. In Springer
International Publishing, Ifmbe Proc, pp. 645-648.
(2015)
M. Augustynek, J. Sterba, J. Cihak, 2014. Hemodynamic
diagnostics shunt for hemodialysis. In Biomedical
Engineering and Environmental Engineering 145, 163
(2014)
J.B. Jeppsson, I. Losell, 1994. Set of tubes for peritoneal
dialysis. U.S. Patent No. 5,338,293. 16 Aug. 1994.
J. Brugger, K. J. Manica, W.G. Palsulich, 1997. Dialysis
blood tubing set. U.S. Patent No. 5,693,008. 2 Dec.
1997.
J. Brugger, 1999. Technique for using a dialysis machine
to disinfect a blood tubing set. U.S. Patent No.
5,948,251. 7 Sep. 1999.
L. Di Landro, C. Capone, F. Inzoli, P.E. Malacari, 2005.
Coextruded PVC tubes for biomedical application. In
Journal of Vinyl and Additive Technology, 11(3), 111-
118.
J. Neergaard, B. Nielsen, V. Faurby, D.H. Christensen,
O.F. Nielsen, 1971. Plasticizers in PVC and the
Occurence of Hefatitis in A Haemodialysis Unit: A
Preliminary Communication. In Scandinavian journal
of urology and nephrology, 5(2), 141-145..
I. Kostić, T. Anđ
elković, D. Anđelković, A. Bojić, T.
Cvetković, D. Pavlović, 2017. Quantification of
DEHP into PVC components of intravenous infusion
containers and peritoneal dialysis set before and after
UV-A treatment. In BULGARIAN CHEMICAL
COMMUNICATIONS, 49(2), 360-365.
N. Beisser, S. Oesterreich, 2018. Heating apparatus for
heating dialysis liquid, dialysis liquid tube set, set,
medical apparatus and methods. U.S. Patent
Application No 15/879,656, 2018.
G. Latini, M. Ferri, F. Chiellini, 2010. Materials
degradation in PVC medical devices, DEHP leaching
and neonatal outcomes. In Current medicinal
chemistry, 17.26 (2010): 2979-2989.
BIODEVICES 2019 - 12th International Conference on Biomedical Electronics and Devices
234