of the elasticity E of the vessel wall and a decrease in
its diameter ๐ท lead to a increase in the value ๐ of
PWV.
The natural method of measuring the pulse wave
travel time (PWTT) between a pair of given points
located proximally on the arterial wall is a direct
method of measuring PWV. Direct measurement of
PWV requires a pair of sphygmometric sensors
located proximally above the superficial vessels
(arteries) and distal to the heart. Any of the pairs of
points located above the carotid, femoral, radial, and
other palpable arteries are suitable for this role. If the
distance ๐ between such pair of points is known and
the delay โ๐ก of pulse wave transition time between
these points is measured, then the PWV value is
obviously the ratio of the first of them to the second:
๐=
๎ฏ
โ๎ฏง
=
๎ฏ
๎ฏง
๎ฐฎ
๎ฌฟ๎ฏง
๎ฐญ
,
(2)
where ๐ก
๎ฌถ
and ๐ก
๎ฌต
are the time moments when the pulse
wave passes through the locations of the selected
points. How these moments are measured is not an
easy question. They can be measured by the event of
passing of a specific waveform-related label, which
can be the foot of the waveform, its maximum, the
maximum slope of the wavefront, etc. (Katsuura
, T. et
al, 2017
).
2 PWV ESTIMATION BASED ON
CONTINUOUS BLOOD
PRESSURE REGISTRATION
Earlier, we developed and tested a new three-channel
pneumatic sensor for continuous non-invasive blood
pressure monitoring (
Antsiperov, 2020). An obvious
advantage of the developed device is the possibility
of a continuous measurement of the dynamics of
blood pressure, which allows not only to determine
the current systolic / diastolic pressure, but also to
track the dynamics of blood pressure, both within the
cycle and at significant time intervals. At the same
time, it is not always possible to correctly calibrate
the measured value in pressure units. For correct
measurement of blood pressure in absolute units, a
certain position of the sensor above the artery is
required, as well as a rigid base below it, such as the
radial bone for the artery of the same name
(
Antsiperov, 2020). Only under the properly positioned
sensor the pressure in the working chamber of the
sensor could be considered equal to the blood
pressure in the artery (Figure. 1)
It was found experimentally that for arteries
whose location does not satisfy the above conditions,
it is possible to observe a pressure pulse wave signal,
the level and amplitude of which is noticeably
distorted by viscoelastic tissues lying both between
the sensor and the artery and beneath artery.
Figure 1: Three-channel pneumatic sensor for continuous
non-invasive arterial blood pressure monitoring (B). The
difference is the shape of the pulse wave signal from the
sensor (A), depending on the position of the sensor pads: โ
โ pad is directly over the artery, โ . โฆ โ pad is shifted to the
left and to the right respectively from the central projection
of the radial artery.
However, the general structure and corresponding
temporal characteristics of the signal are retained in
this case as well. This effect can be used to measure
the delay of signal front relative to the ECG reference
element. To solve the problem of the assessment of
pulse wave propagation velocity a unique way to use
pneumatic sensors was developed. The idea was the
following: if it is possible to take measurements for a
pair of points on the artery for a finite time with the
patient's condition unchanged, then you can try to do
it with the only one sensor. Evidently, the position of
the R-wave on the ECG can be used as a periodically
repeating reference "zero". For this purpose, an
additional channel for synchronous measurement of
the electrocardiographic (ECG) signal was integrated
into the pressure sensor configuration. In this case,
speaking in the language of radio engineering, this
channel acts as a kind of reference signal. The ECG
amplifier circuit was developed with the expectation
of using dry electrodes without a conductive gel and
without a neutral electrode, that required the
application of both analogue circuitry and digital
filters. The simplified scheme was selected to
minimize the inconvenience when applying and
removing the electrodes and is used so far at the
development stage only.