analyzes the effects of melanin and the depth of the
air gap on the ratio of ratios and the accuracy of SpO
2
estimates. The main observations from this work and
future directions are discussed next.
It can be observed from Figure 4 that the slopes of
SpO
2
vs. the ratio of ratios curves are different for
different skin colors. Therefore, if we could estimate
the tone of the subject’s skin, we would be able to
adjust the curves or select the appropriate regression
curve for calibration. In addition, if the SpO
2
calibration curve computed for subjects with a lower
level of melanin is used for subjects with a larger level
of melanin, the results from Figure 5 do not confirm
that the overestimation would occur. This might be
because of outliers of SpO
2
estimates at 0.3% and 8%
melanin levels. The results, however, show variability
with an increased level of melanin.
The illumination is significantly reduced with the
increased level of melanin, as shown in Figure 3. It
drops faster for red than for infrared light. This
information can be used to increase the radiated
power of the LEDs in case a darker skin color is
detected.
In reflectance pulse oximetry, it is difficult and
sometimes impossible to control how well the probe
is attached to the skin. This can result in a varying air
gap. Increasing the air gap results in an increased
level of reflection of light from the surface of the skin,
resulting in less light propagating in the tissue and a
weaker AC component of the signal. Figure 6 shows
changes in illumination fraction at the photodetector
with increasing the air gap and barrier depth. Figure
7 shows changes in the estimated SpO
2
for the same
case. Therefore, the pressure applied to the device
should be controlled and constant to allow for a fixed
air gap. Also, motion artifacts cause varying air gaps;
therefore, they should be detected, and the signal
obtained during motion artifacts should be removed
from the analysis.
The study can be extended in several ways. One is
to repeat measurements for different points and to run
Monte Carlo simulations longer to get less noisy data.
We can also modify the tissue model to more
precisely reflect the body site where the probe is
attached. For example, the optical properties and
thickness of the skin layers on the finger, ear, wrist,
neck, or forehead (which are some of the typical sites)
are different.
The simulated depth of the blood vessels can be
modified based on the selected site. In the simulation,
we assume that the blood volume changes occur only
in the cutaneous plexus. It was also observed
(Mannheimer, 2004) that placing the reflectance
pulse oximetry probe directly over a larger blood
vessel can degrade SpO
2
estimation accuracy. This
can also be modeled.
Future work will also include building an end-to-
end pulse oximeter simulator that will include the
model of the LEDs and driving circuits, the model of
the tissue and light propagation and the model of the
photodiode and the front-end electronics. This will
allow us to understand further how noisy the signal is
in case of small levels of illumination of the
photodetector and to perform sensitivity analysis to
understand what components/parameters of the
system affect the output the most.
REFERENCES
Al-Halawani, R., Chatterjee, S. & Kyriacou P.A. (2022).
Monte Carlo Simulation of the Effect of Human Skin
Melanin in Light-Tissue Interactions. 2022 44th Annual
International Conference of the IEEE Engineering in
Medicine & Biology Society (EMBC).
Arefin, M.S., Dumont, A.P. & Patil, C.A.(2022). Monte
Carlo based simulations of racial bias in pulse oximetry.
Proc. SPIE 11951,Design and Quality for Biomedical
Technologies XV, 1195103.
Bickler, P.E., Feiner, J.R & Severinghaus, J.W., (2005).
Effects of Skin Pigmentation on Pulse Oximeter
Accuracy at Low Saturation, Anesthesiology, 102:
715–9.
Bolic M. (2023). Pervasive Cardiovascular and Respiratory
Monitoring Devices: Model-Based Design. Elsevier.
Cajas, S. A., Landínez, M. A. & López, D. M. (2020)
Modeling of motion artifacts on PPG signals for heart-
monitoring using wearable devices. 15th International
Symposium on Medical Information Processing and
Analysis, 11330(3).
Chatterjee, S. & Phillips, J. P. (2017). Investigating optical
path in reflectance pulse oximetry using a multilayer
Monte Carlo model. Clinical and Preclinical Optical
Diagnostics, edited by J. Quincy Brown, Ton G. van
Leeuwen, Proc. of SPIE-OSA, 10411.
Fuentes-Guajardo, M., Latorre, K., León, D. & Martín-
Escudero, P. (2022). Skin Pigmentation Influence on
Pulse Oximetry Accuracy: A Systematic Review and
Bibliometric Analysis. Sensors, 22(3402).
Hartmann, V., Liu, H., Chen, F., Qiu, Q., Hughes, S. &
Zheng D. (2019). Quantitative Comparison of
Photoplethysmographic Waveform Characteristics:
Effect of Measurement Site. Front Physiol., 10(198).
Mannheimer, P.D., et al. (2004). The Influence of Larger
Subcutaneous Blood Vessels on Pulse Oximetry.
Journal of Clinical Monitoring. 18, 179-88.
Marti, D., Aasbjerg, R. N., Andersen P. E. & Hansen A. K.
(2018). MCmatlab: an open-source, user-friendly,
MATLAB-integrated three-dimensional Monte Carlo
light transport solver with heat diffusion and tissue
damage. Journal of Biomedical Optics, SPIE, 23, 1-6..
BIODEVICES 2023 - 16th International Conference on Biomedical Electronics and Devices