The results of the validity of the ET in Table 1
show that the validity of the data measured in lab
environment was higher with an average validity of
89.62% ± 7.53% than in flight environment with
71.00% ± 10.71%. This is due to the fact, that during
microgravity phases in flight environment the
operator starts floating around. With the given
experimental design, it was not possible to fixate the
operator fully on the plane ground, so the operator
floated out of the region of measurement from the ET.
In lab environment, however, it was not the case since
the operator was sitting on a chair in Earth gravity.
The validity was also not expected to have 100%
because for each eye blink pupil images cannot be
detected. If the validity is compared to the number of
eye blinks (Fig. 4A), the validity is lower for
conditions with a higher number of eye blinks and is
also proportional related to the maximum blink
duration (Fig. 4C), which was expected.
The analysis of the number of eye blinks (cf. Sec.
4.1.1) shows an increase of the number of eye blinks
with an increase of WL condition in both
environments except for BU87D in lab environment
(cf. Fig. 4A). An increase of the number of eye blinks
for increasing WL condition was also expected from
literature (Volden, Alwis, de Viveka, & Fostervold,
2018) and is supported in this study by the higher
validity percentage for lower WL condition (see Tab.
1). However, these results must be viewed with
caution, as the duration of each run was only 22s and
the standard number of blinks per minute are between
2 and 50 (Monster, Chan, & O'Connor, 1978). This
means that if the operator blinks less in general, the
number of blinks over a short period of time is not
representative. As a conclusion, it seems like this
parameter is technically measurable in microgravity,
but within this experimental setup it is very limited
due to the short period of time where it is measured.
The blink frequency was analysed as well (cf. Sec.
4.1.1) and results were contrary to the results of the
number of eye blinks. From literature it was expected
that the blink frequency is increasing with increasing
WL condition (Volden, Alwis, de Viveka, &
Fostervold, 2018), but Figure 4B shows a decrease in
three out of four cases. For the fourth case, BU87D in
lab environment, the median blink frequency
increases, but in most of the runs there were less than
two eye blinks, so it was not possible to calculate the
frequency of eye blinks within these runs, which
means that this result is not representative. As already
discussed before, 22s for each run within this
experiment is not sufficient for analysing changes in
the WL levels of the operators based on ET
parameters. Furthermore, the calculation of the blink
frequency is another limiting factor. Due to long
phases of closed eyes at the beginning of single runs,
as well as inaccuracies of measures due to the onset
of hyper-gravity at the end of each run, the blink
frequency was only calculated in the time between the
first and the last recognised blink. This leads to a low
robustness of the calculation for a small sample size,
which becomes clear due to the large variation of the
calculated frequencies, e.g., for Operator BU87D in
the lab environment (cf. Fig. 4B). Operator AC07D
also shows many outliers in the flight environment,
which indicate this. It seems like this parameter is not
suitable for experimental setups with short-term
measurements independent of the prevailing
gravitational conditions.
The results of the maximum blink duration are
shown in Figure 4C, which illustrates an increase of
the overall median of the maximum blink duration for
higher WL condition in comparison to lower WL
condition in all environments except for BU87D in
lab environment. For BU87D in lab environment the
maximum blink duration decreases. The results are
inhomogeneous which could be due to the small
sample size of only two operators within this study.
Volden et al. (2018) had 21 subjects and removed
four outliers to obtain normal distribution. Also, the
definition of the maximum length of a blink varies. In
this work a blink was defined between 100ms and
500ms (cf. Sec. 3.4) regarding the work of Aksu et al.
(2024), because a typical blink lasts between
100ms – 300ms and blinks about 500ms are
considered as drowsiness (Johns, 2003). Volden et al.
(2018) did not define time limits for eye blinks and
had a mean maximum blink duration of
973.21ms ± 637.66ms, which would be drowsiness
instead of eye blinks regarding the definition within
our work. Therefore, this parameter needs to be more
investigated in Earth gravity as well as in
microgravity with longer periods of measurements
and the given limitations.
The last parameter analysed for ET data was the
pupil diameter for both eyes (Sec. 4.1.2). Figure 4D
depicts that the median pupil diameter over all runs of
the right eye decreases for higher WL condition in
comparison to lower WL condition for both operators
in both environments. For flight environment the
decrease seems higher and could be explained by the
use of scopolamine since ophthalmic adverse effects
are expected with that (Merck & Co., Inc., 2024). The
pupil diameter seems also not to be a suitable
parameter in microgravity with scopolamine intake
and also need further tests for Earth gravity.
The HR was analysed from ECG data with respect
to the WL conditions (Fig. 5A). The HR decreased for