Abstract

Development of bioanalytical assays for determination of glycated hemoglobin content in blood samples is reported. First, a combined biosensor setup for determination of total and glycated hemoglobin content was successfully developed and tested. The effect of various operating parameters, such as ionic strength, flow rate and instrumental set-up was optimized. The total hemoglobin content was analyzed by measuring of absorbance of the hemoglobin-cyanide derivative at 540 nm. Only one standard (calibrator), diluted in various proportions, was necessary for the method calibration. The full range of HbA1c content (4 to 15 %) presented in blood can be analyzed. Only 1 μl of blood was required for analysis. The developed method was successfully evaluated for analysis of blood samples collected from diabetic patients. Next, the heterogeneous affinity assay performed in a microtitre plate with an immobilized boronic acid is described. This assay is based on ELISA (Enzyme-Linked Immunosorbent Assay) principle; however stable chemiselective ligand is used in this case. The content of glycated hemoglobin is determined according to its peroxidase activity after attachment to immobilized boronic acid derivative; the total hemoglobin concentration is measured as an absorbance at 405 nm.

Download


Paper Citation


in Harvard Style

Přibyl J. and Skládal P. (2008). BIOINTERFACES BASED ON IMMOBILIZED BORONIC ACID WITH SPECIFITY TO GLYCATED PROTEINS . In Proceedings of the First International Conference on Biomedical Electronics and Devices - Volume 2: BIODEVICES, (BIOSTEC 2008) ISBN 978-989-8111- 17-3, pages 107-112. DOI: 10.5220/0001051501070112


in Bibtex Style

@conference{biodevices08,
author={Jan Přibyl and Petr Skládal},
title={BIOINTERFACES BASED ON IMMOBILIZED BORONIC ACID WITH SPECIFITY TO GLYCATED PROTEINS},
booktitle={Proceedings of the First International Conference on Biomedical Electronics and Devices - Volume 2: BIODEVICES, (BIOSTEC 2008)},
year={2008},
pages={107-112},
publisher={SciTePress},
organization={INSTICC},
doi={10.5220/0001051501070112},
isbn={978-989-8111- 17-3},
}


in EndNote Style

TY - CONF
JO - Proceedings of the First International Conference on Biomedical Electronics and Devices - Volume 2: BIODEVICES, (BIOSTEC 2008)
TI - BIOINTERFACES BASED ON IMMOBILIZED BORONIC ACID WITH SPECIFITY TO GLYCATED PROTEINS
SN - 978-989-8111- 17-3
AU - Přibyl J.
AU - Skládal P.
PY - 2008
SP - 107
EP - 112
DO - 10.5220/0001051501070112