MICROCOMPUTERIZED SYSTEM TO ASSESS THE PERFORMANCE OF LUNG VENTILATORS
Daniel Marinho Silva, Maurício Campelo Tavares, Raimes Moraes
2009
Abstract
Lung ventilators may harm patients if they are not properly calibrated. Therefore, they must be periodically assessed to verify if the supplied volume and pressure match the ventilatory settings. This paper presents a system based on a microcomputer to assess the performance of lung ventilators. The hardware of the developed system contains three modules: transducer, conditioning and acquisition. The transducer module converts, to electrical signals, the pressure and flow waveforms supplied to a lung simulator by a lung ventilator. It also supplies digital measurements of temperature and relative humidity (RH). The conditioning module amplifies and filters the pressure and flow signals. The acquisition module reads the digital measurements (temperature and RH) and carries out the analog to digital conversion of the conditioning module outputs, sending these data to the microcomputer via radio-frequency. Software written in C++ shows the acquired waveforms on the PC screen and calculates the parameters required by the IEC 60601-2-12. Data on the lung ventilator model, the sampled waveforms and the calculated parameters are stored in a database, allowing the equipment follow-up. Comparative result of tests carried out with the developed system and with commercial equipment is presented.
References
- Analog Devices 2003, ADuC841/ADuC842/ADuC843 microcConverter 12-Bit ADCs and DACs with embedded high speed 62-kB Flash MCU datasheet. USA.
- Doeblin, E. O., 1990. Measurement Systems: application and design. Chapter 7, McGraw-Hill, Singapore.
- Fernández-Pérez, E. R., Keegan, M. T., Brown, D. R.; Hubmayr, R. D., Ognjen G. 2006. Intraoperative Tidal Volume as a Risk Factor for Respiratory Failure after PneumonectomyAnesthesiology, 105: 14-18.
- Honeywell Inc., 2008. DC030NDC datasheet. http://sensing.honeywell.com/index.cfm?ci_id=14030 1&la_id=1&pr_id=82004
- International Electrotechnical Commission, 2001. IEC 60601-2-12: Medical electrical equipment - Part 2-12: Particular requirements for the safety of lung ventilators - Critical care ventilators. Switzerland.
- Jones, R. P., Conway D. H, 2005. The effect of electromagnetic interference from mobile communication on the performance of intensive care ventilator, European Journal of Anaesthesiology; 22: 578-583.
- Kcwirefree, 2006. KC-210 embedded bluetooth USB serial adapter datasheet. USA.
- Kcwirefree, 2007. KC-21 OEM bluetooth module datasheet. USA.
- Pierce, L. N. B., 1995. Guide to Mechanical Ventilation and Intensive Respiratory Care, W. B. Saunders Company, USA.
- Ricard, J. D., Dreyfuss, D., Saumon, G., 2003. Ventilatorinduced lung injury, Eur Respir J, 22: Suppl. 42, 2s9s.
- Sensirion, 2007. SHT1x / SHT7x humidity & temperature sensor datasheet v3.01, Switzerland.
- Tolotti, A. A., 2004. Desenvolvimento de procedimento para realização de ensaio de desempenho em ventiladores pulmonares, Monograph, Clinical Engineering Graduate Course, Federal University of Santa Catarina, Florianópolis, Brazil.
- Wallin, M. K. E. B., Wajntraub S., 2004. Evaluation of Bluetooth as a Replacement for Cables in intensive Care and Surgery, Anesth Analg; 98:763-767.
- Wrigge, H., Uhlig, U., Zinserling, J. , Behrends-Callsen, E., Ottersbach, G., Fischer, M., Uhlig, S., Putensen, C., 2004. The Effects of Different Ventilatory Settings on Pulmonary and Systemic Inflammatory Responses During Major Surgery, Anesthesia & Analgesia, 98: 775-781.
Paper Citation
in Harvard Style
Marinho Silva D., Campelo Tavares M. and Moraes R. (2009). MICROCOMPUTERIZED SYSTEM TO ASSESS THE PERFORMANCE OF LUNG VENTILATORS . In Proceedings of the International Conference on Biomedical Electronics and Devices - Volume 1: BIODEVICES, (BIOSTEC 2009) ISBN 978-989-8111- 64-7, pages 161-166. DOI: 10.5220/0001432101610166
in Bibtex Style
@conference{biodevices09,
author={Daniel Marinho Silva and Maurício Campelo Tavares and Raimes Moraes},
title={MICROCOMPUTERIZED SYSTEM TO ASSESS THE PERFORMANCE OF LUNG VENTILATORS},
booktitle={Proceedings of the International Conference on Biomedical Electronics and Devices - Volume 1: BIODEVICES, (BIOSTEC 2009)},
year={2009},
pages={161-166},
publisher={SciTePress},
organization={INSTICC},
doi={10.5220/0001432101610166},
isbn={978-989-8111- 64-7},
}
in EndNote Style
TY - CONF
JO - Proceedings of the International Conference on Biomedical Electronics and Devices - Volume 1: BIODEVICES, (BIOSTEC 2009)
TI - MICROCOMPUTERIZED SYSTEM TO ASSESS THE PERFORMANCE OF LUNG VENTILATORS
SN - 978-989-8111- 64-7
AU - Marinho Silva D.
AU - Campelo Tavares M.
AU - Moraes R.
PY - 2009
SP - 161
EP - 166
DO - 10.5220/0001432101610166