Abstract

Studies prove, that ECG monitoring over longer periods of time can significantly improve the diagnosis of heart failure. Dry electrodes are a valuable alternative to conventional wet electrodes, if the monitoring period is longer than one week, because they are more stable and cause less skin irritations. One drawback however is the minor signal quality compared to wet electrodes. As there are different electrode materials for dry electrodes available, a qualitative comparison concerning signal quality is carried out in this work. The signal quality of well known dry-electrode-materials used in fitness-heart belts are compared with metal dry electrodes. All electrodes are assembled in an identical mechanical electrode-setup. A parcours-based field trial with 13 subjects is carried out in two stages. In order to estimate the reproducibility of the parcours, the first stage is used to benchmark the differences in signal quality of consecutive measurements not originating from the electrode material. In the second stage however, the actual measurement takes place. The accuracy of the method is presented and the comparison results are discussed.

References

  1. A. Grü tzmann, S. H. and Mü ller, J. (2007). Novel dry electrodes for ecg monitoring. Physiological Measurement, 28.
  2. Baek JY, An JH, C. J. e. a. (2008). Flexible polymeric dry electrodes for the long-term monitoring of ecg. In Sensors and Actuators A: Physical. Elsevier.
  3. Connolly, M. and Buckley, D. A. (2004). Contact dermatitis from propylene glycol in ECG electrodes, complicated by medicament allergy, volume 50 of Contact Dermatitis. Wiley-Blackwell.
  4. David Ruhlemann, Katrin Kuegler, B. M. and Frosch, P. J. (2010). Contact dermatitis to self-adhesive ECG electrodes, volume 62 of Contact Dermatitis. WileyBlackwell.
  5. Fuhrhop, S., Lamparth, S., and et al, K. (2009). Ambulant ecg recording with wet and dry electrodes. In World Congress on Medical Physics and Biomedical Engineering, volume 25/5 of IFMBE Proceedings. Springer.
  6. G. Gargiulo, P. Bifulco, M. C. e. a. (2010). An ultra-high input impedance ecg amplifier for long-term monitoring of athletes. In Medical Devices: Evidence and Research. Dove Press.
  7. Hoffmann, K.-P. and Ruff, R. (2007). Flexible dry surfaceelectrodes for ecg long-term monitoring. In Engineering in Medicine and Biology Society, 2007. EMBS 2007.
  8. M. Di Rienzo, F. Rizzo, G. P. e. a. (2005). Magic system: a new textile-based wearable device for biological signal monitoring. In Engineering in Medicine and Biology Society, 2005. IEEE-EMBS 2005.
  9. PJ. Xu, H. Zhang, X. T. (2008). Textile-structured electrodes for electrocardiogram. Textile Progress.
  10. Pola, T. and Vanhala, J. (2007). Textile electrodes in ecg measurement.
  11. Puurtinen, M., Komulainen, S., Kauppinen, P., Malmivuo, J., and Hyttinen, J. (2006). Measurement of noise and impedance of dry and wet textile electrodes, and textile electrodes with hydrogel. In Engineering in Medicine and Biology Society, 2006. EMBS 7806.
  12. Scilingo, E. P., Gemignani, A., Paradiso, R., Taccini, N., Ghelarducci, B., and De Rossi, D. (2005). Performance evaluation of sensing fabrics for monitoring physiological and biomechanical variables. Information Technology in Biomedicine, IEEE Transactions on, 9(3).
  13. Yoo, J., Yan, L., Lee, S., Kim, H., and Yoo, H.-J. (2009). A wearable ecg acquisition system with compact planarfashionable circuit board-based shirt. Information Technology in Biomedicine, IEEE Transactions on, 13(6).
Download


Paper Citation


in Harvard Style

Fuhrhop S., Lamparth S., Seemann T., Stork W., Heuer S. and Kirst M. (2011). COMPARING DRY ELECTRODE MATERIALS FOR LONG-TERM ECG MONITORING - Evaluation of a New Method . In Proceedings of the International Conference on Bio-inspired Systems and Signal Processing - Volume 1: BIOSIGNALS, (BIOSTEC 2011) ISBN 978-989-8425-35-5, pages 485-488. DOI: 10.5220/0003165104850488


in Bibtex Style

@conference{biosignals11,
author={Silvester Fuhrhop and Stefan Lamparth and Tobias Seemann and Wilhelm Stork and Stephan Heuer and Malte Kirst},
title={COMPARING DRY ELECTRODE MATERIALS FOR LONG-TERM ECG MONITORING - Evaluation of a New Method},
booktitle={Proceedings of the International Conference on Bio-inspired Systems and Signal Processing - Volume 1: BIOSIGNALS, (BIOSTEC 2011)},
year={2011},
pages={485-488},
publisher={SciTePress},
organization={INSTICC},
doi={10.5220/0003165104850488},
isbn={978-989-8425-35-5},
}


in EndNote Style

TY - CONF
JO - Proceedings of the International Conference on Bio-inspired Systems and Signal Processing - Volume 1: BIOSIGNALS, (BIOSTEC 2011)
TI - COMPARING DRY ELECTRODE MATERIALS FOR LONG-TERM ECG MONITORING - Evaluation of a New Method
SN - 978-989-8425-35-5
AU - Fuhrhop S.
AU - Lamparth S.
AU - Seemann T.
AU - Stork W.
AU - Heuer S.
AU - Kirst M.
PY - 2011
SP - 485
EP - 488
DO - 10.5220/0003165104850488