SELECTIVELY BONDED POLYMERIC CHECK VALVE FOR THE RELIABLE REGULATION OF INTRAOCULAR PRESSURE
Seongmin Im, Jaeyong An, Jongchan Choi, Byungphil Mun, Sung Yang, Jonghyun Lee
2011
Abstract
A glaucoma drainage device for the regulation of intraocular pressure is presented. The device consists of three biocompatible polymer layers: the cover (top layer), the diaphragm (intermediate layer), and the baseplate with a cannular (bottom layer). Finite element analyses (FEA) were performed to find optimal design parameters for the targeted cracking pressure: the diameter of orifice, the thickness of intermediate layer, the width of the channel, and the overlapped length of the orifice over valve seat. Top and bottom layers were made by polydimethylsiloxane (PDMS) replica molding process, and an intermediate thin layer with through-holes was fabricated by PDMS fill-in molding method and hydrophobic treatment. The overlapped area of the bottom layer was coated with Cr and Au by using PDMS as a shadow mask. Metal layers are not to be bonded between the bottom and the intermediate layers so that the device showed the enhanced reliability in operation and the higher yield in production. Oxygen plasma treatment was performed for irreversible bonds between separate three PDMS layers. The experimental cracking pressure of the fabricated valve was 2.50 kPa, which is very close to the target value (2.67 kPa). The experiments showed that the proposed polymer check valve can regulate the pressure of the aqueous humour, fluid in an anterior chamber, within the normal intraocular pressure range (15~20mmHg) with a high repeatability.
References
- Kwang W. Oh, Chong H. Ahn, 2006, “A Review of Microvalves”, J. Micromech. Microeng, Vol. 16, pp. 13-39
- Ronalee Lo, Po-Ying Li, Saloomeh Saati, Rajat N. Agrawal, Mark S. Humayun, Ellis Meng, 2009, “A Passive MEMS Drug Delivery Pump for Treatment of Ocular Diseases”, Biomedical Microdevices, DOI 10.1007/s10544-009-9313-9
- Noo Li Jeon, Daniel T. Chiu, Christopher J. Wargo, Hongkai Wu, Insung S. Choi, Janelle R. Anderson, George M. Whitesides, 2002 "Design and Fabrication of Integrated Passive Valves and Pumps for Flexible Polymer 3-Dimensional Microfluidic Systems", Biomedical Microdevices, 4:2,117-121
- Quigley H. A., 1996, Number of people with glaucoma worldwide., Br J Ophthalmol; 80:389-393
Paper Citation
in Harvard Style
Im S., An J., Choi J., Mun B., Yang S. and Lee J. (2011). SELECTIVELY BONDED POLYMERIC CHECK VALVE FOR THE RELIABLE REGULATION OF INTRAOCULAR PRESSURE . In Proceedings of the International Conference on Biomedical Electronics and Devices - Volume 1: BIODEVICES, (BIOSTEC 2011) ISBN 978-989-8425-37-9, pages 234-237. DOI: 10.5220/0003167302340237
in Bibtex Style
@conference{biodevices11,
author={Seongmin Im and Jaeyong An and Jongchan Choi and Byungphil Mun and Sung Yang and Jonghyun Lee},
title={SELECTIVELY BONDED POLYMERIC CHECK VALVE FOR THE RELIABLE REGULATION OF INTRAOCULAR PRESSURE},
booktitle={Proceedings of the International Conference on Biomedical Electronics and Devices - Volume 1: BIODEVICES, (BIOSTEC 2011)},
year={2011},
pages={234-237},
publisher={SciTePress},
organization={INSTICC},
doi={10.5220/0003167302340237},
isbn={978-989-8425-37-9},
}
in EndNote Style
TY - CONF
JO - Proceedings of the International Conference on Biomedical Electronics and Devices - Volume 1: BIODEVICES, (BIOSTEC 2011)
TI - SELECTIVELY BONDED POLYMERIC CHECK VALVE FOR THE RELIABLE REGULATION OF INTRAOCULAR PRESSURE
SN - 978-989-8425-37-9
AU - Im S.
AU - An J.
AU - Choi J.
AU - Mun B.
AU - Yang S.
AU - Lee J.
PY - 2011
SP - 234
EP - 237
DO - 10.5220/0003167302340237