The aim of this study was to develop a device that measures both oxygen saturation (SpO2) and Blood Pressure (BP), non-invasively, continuously and without a cuff. The pulse oximeter was developed for transmission and reflection mode, for use in finger and wrist, respectively. The oxygen saturation results obtained with the developed device placed on the wrist reveled an high relation with an market device. For BP measurement there were tested 3 different combination of sensors localizations to obtain Pulse Transit Time (PTT): ECG and pulse oximeter on the finger, ECG and pulse oximeter on the wrist and pulse oximeter on the wrist and photoplethismogram on finger, and two calibration modes: for each subject - without input parameters - and general - with one input parameter (height). The results reveled better correlation between the BP estimated with the first calibration and the reference method (cuff-based) than for the second calibration.
References
- Daminani, F. R. (2010). Proposta de protótipo de um oxímetro de pulso empregando tecnologia FPAA. Master's thesis, Universidade de Sa˜o Paulo - Escola de Engenharia de Sa˜o Carlos, So Paulo.
- Fung, P., Dumont, G., Ries, C., Mott, C., and Ansermino, M. (2004). Continuous noninvasive blood pressure measurement by pulse transit time. In Engineering in Medicine and Biology Society, 2004. IEMBS'04. 26th Annual International Conference of the IEEE, volume 1, pages 738-741. IEEE.
- Gesche, H., Grosskurth, D., Kuchler, G., and Patzak, A. (2011). Continuous blood pressure measurement by using the pulse transit time: comparison to a cuffbased method. European Journal of Applied Physiology, pages 1-7.
- Haahr, R. G. (2006). Reflectance Pulse Oximetry Sensor for the Electronic Patch. Master's thesis, Technical University of Denmark - MIC - Department of Micro and Nanotechnology, Denmark.
- Naschitz, J., Bezobchuk, S., Mussafia-Priselac, R., Sundick, S., Dreyfuss, D., Khorshidi, I., Karidis, A., Manor, H., Nagar, M., Peck, E., et al. (2004). Pulse transit time by r-wave-gated infrared photoplethysmography: Review of the literature and personal experience. Journal of clinical monitoring and computing, 18(5):333-342.
- Nonin (2011). Nonin - nonin medical inc. [online] avalilable at: http://www.nonin.com/ [accessed 5 september 2011].
- Payne, R., Symeonides, C., Webb, D., and Maxwell, S. (2006). Pulse transit time measured from the ecg: an unreliable marker of beat-to-beat blood pressure. Journal of Applied Physiology, 100(1):136.
- PS (2011). Portal da saude - ministerio da saude [online] avalilable at: http://www.portaldasaude.pt/ [accessed 5 september 2011].
- Reddy, K., George, B., Mohan, N., and Kumar, V. (2009). A Novel Calibration-Free Method of Measurement of Oxygen Saturation in Arterial Blood. Instrumentation and Measurement, IEEE Transactions on, 58(5):1699-1705.
- Reisner, A., Shaltis, P., McCombie, D., and Asada, H. (2008). Utility of the photoplethysmogram in circulatory monitoring. Anesthesiology, 108(5):950.
- Sola, A. et al. (2005). Pulse oximetry in neonatal care in 2005. A comprehensive state of the art review. In Anales de Pediatría, volume 62, page 266.
- WelchAllyn (2011). Welch allyn [online] avalilable at: http://www.welchallyn.com/[accessed 5 september 2011].
- WHO (2011). World health organization - the top 10 causes of death [online] - available at: http://www.who.int/mediacentre/factsheets/fs310 /en/index.html [accessed 5 september 2011].
- Wong, M., Poon, C., and Zhang, Y. (2009). An evaluation of the cuffless blood pressure estimation based on pulse transit time technique: a half year study on normotensive subjects. Cardiovascular Engineering, 9(1):32-38.
Paper Citation
in Harvard Style
Fé A., Sousa J. and Gamboa H. (2012). DEVELOPMENT OF A PULSE OXIMETER AND BLOOD PRESSURE MEASUREMENT DEVICE . In Proceedings of the International Conference on Biomedical Electronics and Devices - Volume 1: BIODEVICES, (BIOSTEC 2012) ISBN 978-989-8425-91-1, pages 201-206. DOI: 10.5220/0003781402010206
in Bibtex Style
@conference{biodevices12,
author={Ana Fé and Joana Sousa and Hugo Gamboa},
title={DEVELOPMENT OF A PULSE OXIMETER AND BLOOD PRESSURE MEASUREMENT DEVICE},
booktitle={Proceedings of the International Conference on Biomedical Electronics and Devices - Volume 1: BIODEVICES, (BIOSTEC 2012)},
year={2012},
pages={201-206},
publisher={SciTePress},
organization={INSTICC},
doi={10.5220/0003781402010206},
isbn={978-989-8425-91-1},
}
in EndNote Style
TY - CONF
JO - Proceedings of the International Conference on Biomedical Electronics and Devices - Volume 1: BIODEVICES, (BIOSTEC 2012)
TI - DEVELOPMENT OF A PULSE OXIMETER AND BLOOD PRESSURE MEASUREMENT DEVICE
SN - 978-989-8425-91-1
AU - Fé A.
AU - Sousa J.
AU - Gamboa H.
PY - 2012
SP - 201
EP - 206
DO - 10.5220/0003781402010206