Impact of Airflow Rate on Amplitude and Regional Distribution of Normal Lung Sounds

Elmar Messner, Martin Hagmüller, Paul Swatek, Freyja-Maria Smolle-Jüttner, Franz Pernkopf

2017

Abstract

In computerized lung sound research, the usage of a pneumotachograph, defining the phase of respiration and airflow velocity, is essential. To obviate its need, the influence of airflow rate on the characteristics of lung sounds is of great interest. Therefore, we investigate its effect on amplitude and regional distribution of normal lung sounds. We record lung sounds on the posterior chest of four lung-healthy male subjects in supine position with a 16-channel lung sound recording device at different airflow rates. We use acoustic thoracic images to discuss the influence of airflow rate on the regional distribution. At each airflow rate, we observe louder lung sounds over the left hemithorax and a constant regional distribution above an airflow rate of 0.7 l/s. Furthermore, we observe a linear relationship between the airflow rate and the amplitude of lung sounds.

References

  1. Charleston-Villalobos, S., Cortés-Rubiano, S., GonzálezCamerena, R., Chi-Lem, G., and Aljama-Corrales, T. (2004). Respiratory acoustic thoracic imaging (RATHI): assessing deterministic interpolation techniques. Medical and Biological Engineering and Computing, 42(5):618-626.
  2. Fiz, J. A., Gnitecki, J., Kraman, S. S., Wodicka, G. R., and Pasterkamp, H. (2008). Effect of body position on lung sounds in healthy young men. CHEST Journal, 133(3):729-736.
  3. Gavriely, N. and Cugell, D. W. (1996). Airflow effects on amplitude and spectral content of normal breath sounds. Journal of applied physiology, 80(1):5-13.
  4. Hossain, I. and Moussavi, Z. (2002). Relationship between airflow and normal lung sounds. In Proceedings of the 15th IEEE Canadian Conference on Electrical and Computer Engineering (CCECE'02), pages 1120-1122.
  5. Kraman, S. (1984). The relationship between airflow and lung sound amplitude in normal subjects. CHEST Journal, 86(2):225-229.
  6. Messner, E., Hagmüller, M., Swatek, P., and Pernkopf, F. (2016). A robust multichannel lung sound recording device. In Proceedings of the 9th Annual International Conference on Biomedical Electronics and Devices (BIODEVICES'16), pages 34-39.
  7. Murphy, R. (2007). Development of acoustic instruments for diagnosis and management of medical conditions. Engineering in Medicine and Biology Magazine, 26(1):16-19.
  8. Pasterkamp, H., Kraman, S., DeFrain, P., and Wodicka, G. (1993). Measurement of respiratory acoustical signals. Comparison of sensors. CHEST Journal, 104(5):1518-1525.
  9. Sen, I. and Kahya, Y. (2006). A multi-channel device for respiratory sound data acquisition and transient detection. In Proceedings of the 27th Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBS'06), pages 6658-6661.
  10. Shykoff, B. E., Ploysongsang, Y., and Chang, H. (1988). Airflow and normal lung sounds1. 2. Am Rev Respir Dis, 137:872-876.
  11. Torres-Jimenez, A., Charleston-Villalobos, S., GonzalezCamarena, R., Chi-Lem, G., and Aljama-Corrales, T. (2008a). Asymmetry in lung sound intensities detected by respiratory acoustic thoracic imaging (RATHI) and clinical pulmonary auscultation. In Proceedings of the 30th Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBS'08), pages 4797-4800.
  12. Torres-Jimenez, A., Charleston-Villalobos, S., GonzalezCamarena, R., Chi-Lem, G., and Aljama-Corrales, T. (2008b). Respiratory acoustic thoracic imaging (RATHI): Assessing intrasubject variability. In Proceedings of the 30th Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBS'08), pages 4793-4796.
  13. Yosef, M., Langer, R., Lev, S., and Glickman, Y. A. (2009). Effect of airflow rate on vibration response imaging in normal lungs. The open respiratory medicine journal, 3(1).
Download


Paper Citation


in Harvard Style

Messner E., Hagmüller M., Swatek P., Smolle-Jüttner F. and Pernkopf F. (2017). Impact of Airflow Rate on Amplitude and Regional Distribution of Normal Lung Sounds . In Proceedings of the 10th International Joint Conference on Biomedical Engineering Systems and Technologies - Volume 4: BIOSIGNALS, (BIOSTEC 2017) ISBN 978-989-758-212-7, pages 49-53. DOI: 10.5220/0006134400490053


in Bibtex Style

@conference{biosignals17,
author={Elmar Messner and Martin Hagmüller and Paul Swatek and Freyja-Maria Smolle-Jüttner and Franz Pernkopf},
title={Impact of Airflow Rate on Amplitude and Regional Distribution of Normal Lung Sounds},
booktitle={Proceedings of the 10th International Joint Conference on Biomedical Engineering Systems and Technologies - Volume 4: BIOSIGNALS, (BIOSTEC 2017)},
year={2017},
pages={49-53},
publisher={SciTePress},
organization={INSTICC},
doi={10.5220/0006134400490053},
isbn={978-989-758-212-7},
}


in EndNote Style

TY - CONF
JO - Proceedings of the 10th International Joint Conference on Biomedical Engineering Systems and Technologies - Volume 4: BIOSIGNALS, (BIOSTEC 2017)
TI - Impact of Airflow Rate on Amplitude and Regional Distribution of Normal Lung Sounds
SN - 978-989-758-212-7
AU - Messner E.
AU - Hagmüller M.
AU - Swatek P.
AU - Smolle-Jüttner F.
AU - Pernkopf F.
PY - 2017
SP - 49
EP - 53